22

Sequence for a Complete New Setup

Then presuming the system is correctly installed and cabled the process would be as follows:

If you need to perform a complete new set up

1. Use the FIND command to discover connected devices and install them to the controller

2. Set up the required alarm levels and relay actions

3. Test using zero and calibration gases

The following dialogues describe each function to use 

The FIND Command

find

1 of 11

find

1 of 11

find

Release

s=2 40%

checking

1 = 4101
found= 2

2 = 4101
found= 2

2 = 4101

Release

no

Release

yes

Release

saved=2

passed

As previously described enter password mode and enter password 50 to gain access to the 

engineers menu. The first menu option (menu option 1 of 11) is the FIND menu. To run this option 

the detectors must be correctly connected to the controller and displaying green power LED function 

as a minimum (some of the green power LED’s may be flashing if detectors already have 

communication.) The FIND function then works in the following manner

Press and hold button 
until prompted to 
release. The system 
starts discovering 
connected devices 

Display indicates devices found so far

Display indicates % progress 

When the discovery process is 
complete the display indicates 
how many sensors have been 
found. The system also indicates 
the first found address. Each 
button press then indicates each 
found device address in turn.

To store or abort the data press and 
hold the button until prompted to 
release. The first option is NO to 
store data, rotating the jog wheel 
changes this to YES to store data 
and vice versa. With the desired 
option displayed press the select 
button.

If the YES option to store data 
was selected then the number 
of saved devices is shown on 
screen and passed of fail to 
indicate the status of the 
update

1

2

3

4

Relay Test

If the control panel has been shipped pre-configured then once correctly connected the system will be 

operational. The controller should correctly cycle through each channel with no indicated errors. 

The alarm relay outputs can now be connected (if they are being used). The relay outputs can be 

forced on and off using the ‘test relay’ function (TEST RLY).

test rly
13 of 12

Содержание TOC-650-150

Страница 1: ...r a t i o n a n d M a i n t e n a n c e V 3 1 Triton House Crosby Street Stockport Sk2 6SH England T el 44 0 161 483 1415 Fax 44 0 161 484 2345 Email sales internationalgasdetectors co uk Website www...

Страница 2: ...10 4 14 5 7 12 23 27 27 Addressable Detector Node Highway Hub Controller Overview Relay Card Overview Relay Card Setup User Actions Overview Menu overview Putting Into Service New Setup Sequence The S...

Страница 3: ...aster Salve Brief Introduction Jog Wheel Interface Service Maintenance The Control Panel leaves the factory configured for the supply voltage stated on the customers order Should the control panel be...

Страница 4: ...0V AC 50 60Hz by selection 1 LEL 2 x 8 LCD Display and Jog Wheel 0 1ppm Full EC Declaration appears on Page 4 0 to 55 Degrees Centigrade IP ratings do not imply that the equipment will detect gas duri...

Страница 5: ...aratus using software and or digital technologies Excluding requirements for SIL EMC Generic standards Immunity for industrial environments Electrical apparatus for the detection and measurement of fl...

Страница 6: ...which supplies gas at 0 5L Min Important Notes Panel Options TOCSIN 750 Series Controller TOCSIN 750 Series Annunciator Options With Display Relay Output Digital or Analogue Input Flammable Gas Detect...

Страница 7: ...tem cables must be screened refer to installers guide General Power can be taken from system modules for small current devices such as LED beacon sounders and is limited to 100mA per module Highways a...

Страница 8: ...this document prior to installation 2 WIRE SYSTEMS 2 W i r e G a s D e t e c t i o n S y s t e m I n s t a l l e r s G u i d e V x x OLIVER IGD Triton House Crosby Street Stockport SK2 6SH England T...

Страница 9: ...upporting 32 Devices 128 Devices in Total Two display options either 650 style with jog wheel or HMI Style touch screen display Note both displays can be tted at the same time which can allow the HMI...

Страница 10: ...DOWN BUTTON UP BUTTON Each Addressable relay card must have its own unique base address This can be set on the card and is indicated below The relays are then addresses numbered from that base address...

Страница 11: ...this one Base Address set up LED s and interface for the relay card In all Cases Relay contact ratings 5A 250V AC Non Inductive 5A 30V DC Non Inductive Spike suppression must be fitted Note that FAULT...

Страница 12: ...nalogue and digital inputs that can be used to read data onto the system from external devices Setup of the inputs type range addressing etc is controlled via IGD s Android Apps Wiring options are ind...

Страница 13: ...put DVM Test Point DVM Test Point Housing Sealing Environment Temperature Voltage Communication Display Analogue I P Sounder Digital Input Relay Pellistor Port Digital Output Comm Port 2 3 4 5 6 7 1 S...

Страница 14: ...threshold it will not be possible to reset the alarm Once fully installed the TOC 650 controller will continuously monitor connected gas detectors and sensors and compare current values with any set a...

Страница 15: ...D Red Continuous on Alarm Level 2 Flashes on Alarm Level 1 Power LED Green Selector Button Selector Jog Wheel WARMUP 600s o Indicate board tested status end user should see PASSED with a green backlig...

Страница 16: ...ement With the correct password displayed press the centre selection button to go to the menu or function Current Channel Reading Inhibit the control panel for up to 60 minutes Use this option if cali...

Страница 17: ...Use this option to FIND Relay Outputs Use to Access Device Diagnostics Set the System Time Set the System Date Move to the USER 100 Menu Use to enable Disable Reset Bluetooth Use to Energise De energ...

Страница 18: ...8 ENG MENU 7 of 8 EXIT 8 of 8 Production Menu Options Provides access for control panel set up and diagnostic functions Use this option to Reset the Hubcard System Re boot Reset Hubcard to defaults Di...

Страница 19: ...Menu CALibrate Note IGD calibration and gas adaptor kits are available and necessary to perform calibration and bump testing Record all results for traceability find sen 1 of 12 find rly 1 of 12 zero...

Страница 20: ...for detectors and I O interface nodes and the relevant check points Warm Up Period In normal operation mode the TOC 650 communicates to each detector or node in turn and displays the data on screen Pr...

Страница 21: ...ed Check the shipping TOC 650 SET UP REPORT to check how the controller and interfacing detectors and nodes have been configured Ensure that the detector addresses match the document and required site...

Страница 22: ...he following manner Press and hold button until prompted to release The system starts discovering connected devices Display indicates devices found so far Display indicates progress When the discovery...

Страница 23: ...ro gas and zero the channel observe and record result Notes All gas detectors will require periodic ZERO and CALIBRATION The calibration interval depends on a number of environmental factors such as t...

Страница 24: ...e is no gas present by using a portable detector B Use a suitable ZERO gas as follows Zero gas should have a humidity between 0 90 RH O2 CO2 NITROGEN PELLISTOR INSTRUMENT AIR TOXIC GASES NITROGEN The...

Страница 25: ...erate This type of alarm is typically used in safety applications for Oxygen deficiency monitoring where you are monitoring for a falling Oxygen level Press and hold until prompted to release to selec...

Страница 26: ...t a zero or calibration the controller sends the request to the detector head for action and monitors the result This means that detectors can be supplied pre calibrated The reading is now displayed s...

Страница 27: ...Calibrate zero point Standard accessories Supplied 1 x Terminator Set 5 off 1 x Diode Pack 1 x Snubber 1 x Resistor Pack Battery Care Installation The batteries will come pre installed but will be di...

Страница 28: ...ter Disk 5110101 Remote Gassing Port 5134601 40mm Stop Button 5138601 Collector Cone Ring Lock 5138701 T103 Splash Guard 5138801 Collector cone kit MK6 TOC GLAND M20 Snap fit cable glad 5925801 Univer...

Отзывы: