Rev. 1.0
Page 21
START
STOP
The AF Counter
measures the
CTCSS code to 1 Hz
resolution. For
greater resolution,
this meter can be
expanded.
The Modulation meter
and Audio Frequency
Counter have been
configured for a 300 Hz
Low Pass filter for this
test.
Test TX Squelch (CTCSS)
Setup:
1. This COM-120B has been pre-configured for common transmitter test
setups.
2. Press the
key to access the COM-120B System Setup
Utility software.
3. The menu to the right will appear.
4. Select
STEP 3
by positioning the
cursor and pressing
F1
Execute
.
5. The COM-120B will configure itself
for testing Transmitter Squelch or
SUB Audible deviation.
6. Configure the radio to channel
3
, key the radio and observe the
CTCSS tone on the Scope and Deviation meter.
7. The CTCSS tone frequency is measured and displayed by the
AF
FREQUENCY
counter.
8.
To increase the resolution of the
Audio Counter reading, zoom in
reading by
pressing [TAB] [1]
[4] [ENTER]
and press the
F1
key to zoom in on the Audio
Counter.