24
10.2 Reading the set parameter value
1� Change from the RUN mode to the main menu
[●]
2� Select the requested parameter
[
▲
] or [
▼
]
3� Change to the setting mode
> The currently set value is displayed for 30 s�
By pressing [●] several times, the display switch
-
es between parameter and parameter value�
[●]
4� Return to the RUN mode without parameter change
10.3 Error indications / self-diagnostics
Warning message
[SC1]
Short circuit in OUT1�
LED1 for OUT1 flashing
(→ 7 Operating and display elements)
[OL]
Detection zone of volumetric flow or temperature exceeded�
Measured value between 120 % and 130 % of the final value of the measuring
range�
[UL]
Below the detection zone of temperature�
Measured value between -120 % and -130 % of the final value of the measur-
ing range�
[Err]
Unit faulty / malfunction�
Measured value greater than 130 % of the final value of the measuring range�
[C�Loc]
Setting pushbuttons locked, parameter change rejected�
Active IO-Link communication�
[S�Loc]
Setting pushbuttons locked, parameter change rejected�
Unlock using parameter setting software�
[IOE�n]
Malfunctioning� The unit is faulty and must be replaced�
11 Technical data
Technical data and scale drawing at www�ifm�com�