Fnr Type of fault
drEd = OFF
drEd = On
drEd = Ondr
DA
AA
DA
AA
DA
AA
59 Both sensor elements faulty�
A
F
A
F
A
F
61 Interruption measuring element 2 (Pt
1000).
A
F
W
T/B
W
T/B
62 Short circuit measuring element 2 (Pt
1000).
A
F
W
T/B
W
T/B
71 Detected sensor drift exceeds warn-
ing level�
W
T
W
T
W
T
72 Detected sensor drift exceeds alarm
level�
A
X
A
X
A
F
91 Supply voltage outside the operating
voltage range�
W
T
W
T
W
T
92 Operating temperature of the elec-
tronics outside the specified range�
W
T
W
T
W
T
• Diagnostic states of the diagnostic output (DA)
-N: nominal operation, no faults detected�
-W: warning: anomaly detected, operation with restricted accuracy or functionality possible�
-A: alarm: fault detected, reliable operation not possible any more� The output signal of the
diagnostic output depends on the setting of the parameter dOU�
-
X: undefined (a defined reaction of the output cannot be guaranteed in case of a fault).
• Signals of the analogue output (AA)
-T: temperature-proportional current signal�
-T/B: temperature-proportional current signal during backup operation�
-
F: fault value (defined by parameter FOU). As soon as no reliable temperature measure
-
ment is possible any more, a fault level is provided in all cases�
-
X: undefined (a defined reaction of the output cannot be guaranteed in case of a fault).