Manual adjustment to maximum flow (HI-Teach) /
monitoring excess flow
The unit detects the current flow and sets this value as the maximum
value for the LED display (LED 9).
In addition the
position of the display window
within the detec-
tion range can be defined: Shift the LED for the maximum display
value to position 8, 7, 6 or 5. In the case of maximum flow all LEDs
from 0 up to this LED are lit. The LEDs above the range signal excess
flow. If the switch point is above this range, the unit switches in the
case of excess flow.
34
2
1
3
Press the
Mode/Enter
button
three times..
LED 9 flashes.
Press the
Learn/Set
button and
keep it pressed,
after 5 s the LED bar (green) fills
from left to right
(release the button as soon as the
first LEDs light),
LED 9 flashes for approx. 15s,
after this LED 9 is lit.
LO
HI
LO
HI
3 x
M
S
M
S
4
Press the
Learn/Set
button
several times until the requested
LED lights (LED 8, 7, 6 or 5).
Each time the button is pressed
the LED moves back by one position.
When it is lower than LED 5 the
cycle starts again at LED 9.
LO
HI
M
S
5
Press the
Mode/Enter
button
briefly (acknowledgement).
The indication goes off briefly.
The unit stores the current flow as
maximum flow and passes into
the operating mode.
HI
LO
M
S
>5s
Apply the operating voltage. After approx. 15 s the unit is ready.
Allow the medium to flow through the system at the required
maximum flow rate
.
LO
HI
M