SPECIFICATIONS
PRS-300 Series
5
FIGURE 2.1 Typical OPERATING GUIDE Af
fi
xed to Unit
(Please see label af
fi
xed to your unit)
PRS-300 PROGRAMMABLE DECADE RESISTANCE SUBSTITUTER
CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION
MODEL: PRS-300 SN: B1-17341127
PRS-300 Nov. 2016
WARNING
Observe all safety rules when working with high voltages or line voltages. Connect the (GND) terminal to earth ground in order to maintain the case at a safe voltage.
Whenever hazardous voltages (>45 V) are used, take all measures to avoid accidental contact with any live components: a) Use maximum insulation and minimize the use
of bare conductors. b) Remove power when adjusting switches. c) Post warning signs and keep personnel safely away.
IET LABS, INC.
www.ietlabs.com
CAGE CODE: 62015
• Long Island, New York
• Tel: (516) 334-5959 • Email: [email protected]
Accuracy:
±(0.007% + 1 m
Ω
) 2 and 4 Terminal at 23°C ±5°C
Minimum setting:
0.1
Ω
Resolution:
1
μΩ
or 6 digits
Range:
0.1
Ω
- 20 M
Ω
Thermal emf:
< 15
μ
V
Maximum Load:
2 A, 200 V (peak), 0.5 W whichever applies
fi
rst
Resistors:
Precision wire-wound and metal foil
RTD Tables:
9 RTD tables can be entered into memory to allow user selection of temperature
and the correct value of resistance will automatically be programmed.
PT-100 and PT-1000 tables for both Fahrenheit and Celsius are programmed
into memory locations 1 to 4.
Adjustment:
Automatic adjustment procedure utilizing a high precision DMM eliminates the
requirement for manual trimming of resistors.
Terminals
:
Four low-emf, gold-plated, tellurium-cop per 5-way bind ing posts are used for
HI
and
LO
ter mi nal pairs for
CUR RENT
and
SENSE
.
GND
bind ing post is con nect ed
to the case, to the chassis ground and to the earth ground.
Switching time:
<1 second per change
Power requirements:
90 - 264 Vac , 47 - 63 Hz. 30 Watts Max.
Remote Control:
USB:
Standard USB Type B connector on rear panel, uses FTDI driver
Remote Control Options:
IEEE:
GPIB standard 24 pin connector, conforms to IEEE-488.2; SCPI 1994.0
command set; Hardware or software con
fi
gurable addressing range of 1 to 30.
E t h e r n e t :
I E E E 8 0 2 . 3 c o m p l i a n t , S p e e d s 1 0 B a s e T
(10 Mb/s) and 100 BaseT (100 Mb/s), IP Address Static or DHCP, Factory setting
192.168.0.254 static
Содержание PRS-300 Series
Страница 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC...
Страница 16: ...SPECIFICATIONS PRS 300 Series 6...
Страница 24: ...PRS 300 Series IEEE INTERFACE OPTION 14...
Страница 34: ...PROGRAMMING PRS 300 Series 24...
Страница 42: ...PRS 300 Series PRS DMM SOFTWARE 32...
Страница 50: ...PRS 300 Series APPENDIX B 40...