VI-700 V&l Source
2
For convenience to the user, the pertinent specifica-
tions are given in an
OPERATING GUIDE,
shown in Figure 2. 1, affixed to the case of the instru-
ment.
2.1 Voltage - Standard
Output Ranges:
200 mV, 2 V, 20 V, (200 V with HV Option).
Resolution:
100
µ
V, 1 mV, 10 mV,
(100 mV with HV Option).
Accuracy:
±(75 ppm + 50 µV), 200 mV range;
±(75 ppm + 0.25 mV), 2 V range;
±(75 ppm + 2.5 mV), 20 V range;
at 23°C, no load, after warmup.
Output Current:
70 mA (200 mA with HP Option), maximum load,
with LED overload indicator, for conformance
with accuracy specifications; 50-100% current
overload output is available with slightly reduced
accuracy.
Output Impedance:
5 m
Ω
typical, 40 m
Ω
for 20 V range.
Noise
and Ripple:
50
µ
Vrms on 200 mV and 2 V ranges, 100
µ
Vrms
on 20 V range; exclusive of random transients.
Temperature Coefficient:
10 ppm/ºC typical, 15 ppm/º C maximum.
2.2
HP - High Power Option - 200 mA
Load Current Capacity
Output Current:
200 mA, maximum load, with LED
overload indicator, on all ranges.
SPECIFICATIONS
Chapter 2
SPECIFICATIONS
Since significant current may be drawn from this unit,
it is advisable to leave the
mA
terminals open. The
current
OVLD
LED will remain on if the
mA
termi-
nals are open, and this indicator may be disregarded.
The
OVLD
indicator will be off and will behave nor-
mally under normal operation.
2.3 HV - High Voltage Option - 200 V
Output
Output Voltage:
0 to 200 V.
Resolution:
1 mV, 10 mV, 100 mV.
Output Current:
10 mA (may be increased).
Accuracy:
±(0.01% + 0.5 LSD); applies to >20 V.
2.4 Current -Standard
Output Ranges:
200
µ
A, 2 mA, 20 mA.
Resolution:
0. 1
µ
A, 1
µ
A, 10
µ
A.
Accuracy:
±(75 ppm + .05 µA); 200 µA;
±(75 ppm + 0.25 µA), 2 mA;
±(75 ppm + 2.5 µA), 20 mA;
at 23°C, no load, after warmup.
Compliance:
0-20 V; with LED overload indicator.