PRS-370 Series
APPENDIX B
37
COMMON Commands:
*IDN? Returns instrument identi
fi
cation “IET Labs Inc.,PRS-370,<Serial
number>,<software version>”
Serial number in the form “Ax-xxxxxxxx” x = 0 to 9. A = Alpha can be A to Z.
*TST? Self-Test Returns ASCII 0 or 1, 0 = bad or 1 = good
*RST
Returns PRS-370 to default “power on” condition
*SAV <number> Saves current resistance value into memory location 0 to 9
Number is 0 to 9
*RCL< number> Recalls resistance value from memory location 0 to 9
Number is 0 to 9
SYSTEM Commands
SYSTem:SERIal <Ax-xxxxxxxx> Sets the serial number in non-volatile memory < serial num-
ber>
Serial number in the form “Ax-xxxxxxxx” x = 0 to 9. A = Alpha can be A to Z.
SYSTem:SERIal?
Returns Serial number
SYSTem:CONTrast <number> Sets system contrast speci
fi
c number from 0 to 15
SYSTem:KEY <number>
Same as pressing key on front panel of PRS-370
<Number> is 0 to 24
keys 0-9
0 – 9
OPEN
14
MENU
15
TABLE
16
Decimal
Point
21
BACK
23
ENTER
24
Short
17
Open
18
Appendix B
PRS-370 COMMON COMMANDS