PRS-370 Series

OPERATION

9

guide.

There are no user serviceable parts inside the 
PRS-370.  Servicing should be referred to qual-
i

fi

 ed personnel.  

3.1.9 Cleaning

To prevent shock unplug the 1863 and 1864 from 
mains prior to cleaning.

Use a dry cloth or a cloth slightly dampened with 
water to clean the external case parts. 

Do not use detergents or chemical solvents. 

Do not attempt to clean internally.

Denatured alcohol can be used to clean insula-

tors around binding posts.

3.2 Connection

3.2.1 General Considerations

The PRS-370 Programmable Decade Resistance 
Substituter utilizes a 5-terminal connection. The 
binding posts are standard laboratory type and 
readily accept banana plugs, spade lugs, alligator 
clips, and bare wire. Binding posts are located on 
the front panel and rear panel of the instrument.  

The Kelvin terminals consisting of a 

CURRENT

 

and a 

SENSE

 pair, each labeled 

HI

 and 

LO

.  

These minimize contact resistance. 

4 Wire

 

connection 

The gold shorting links must be removed on the 
binding posts of the PRS-370 when set to 4 Wire.

2 Wire connection 

The gold shorting links must be used on the 

binding posts of the PRS-370 when set to 2 Wire.  
See section 3.4.4 for more information

The 

GND

 terminal on all models is connected to 

the case and to earth and chassis grounds. This 
may be used as a shield terminal.

3.2.2 Electrical Considerations

The performance of the PRS-370 is directly 
affected by the quality of the connection to 
the system under test. This is particularly true 
with the precision series models having higher-
accuracy and/or lower-impedance decades.

For optimum performance, contact resistance 
should be kept to a minimum by using the most 
substantial mating connection possible, and by 
assuring that the connection is well secured to 
the binding posts. 

3.3 Condensed Operating Instructions

1.  Turn on the PRS-370 ON using POWER 
SWITCH
2.  From the Home screen, select a desired 
resistance value in one of the following ways:

a)  Enter a value using the number keys, then 
press 

ENTER

.

b)  Press the resistance value shown on the 
touch screen and then enter the resistance 
value using the touch screen or keypad.

3.  The source resistance will be output to the 
front or rear binding posts and is indicated at on 
the bottom half of the Home screen.  Front and 
rear binding posts can be selected by going to 
Main Menu>Resistance Settings.  
4.  2-Wire or 4-Wire connection mode is shown 
at the top of the Home screen.  This can be 
changed by going to Main Menu>Resistance 
Settings.

Содержание PRS-370 SERIES

Страница 1: ...NSTRUMENTS FOR TEST AND MEASUREMENT TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC PRS 370 SERIES Programmable Decade Resistance Substituter IEEE 488 2 Ethernet and USB Operation Manual ...

Страница 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC ...

Страница 3: ...EASUREMENT TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC Copyright 2019 IET Labs Inc PRS 370 IM March 2019 PRS 370 SERIES Programmable Decade Resistance Substituter IEEE 488 2 Ethernet and USB Operation Manual ...

Страница 4: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT ...

Страница 5: ... at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warranty IET shall not be liable for any indirect special or consequential damages even if notice has been given to the possibility of such damages THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIED...

Страница 6: ...or it is not marked with a measurement category its measurement terminals must not be connected to line voltage mains The PRS 370 is an indoor use product DANGEROUS PROCEDURE WARNINGS Comply with all WARNINGS Procedures throughout in this manual and instructions on the instrument prevent you from potential hazard These instructions contained in the warnings must be followed BEFORE APPLYING POWER V...

Страница 7: ... be made by qualified maintenance personnel only DO NOT substitute parts or modify the instrument When working with high voltages post warning signs train personnel and keep unauthorized personnel away Do not apply any voltage or currents to the terminals of the instrument in excess of the maximum limits indicated in the specifications section of this manual To avoid the danger of introducing addi...

Страница 8: ...2002 96 EC This product complies with the WEEE Directive 2002 96 EC marking requirements The affixed label indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE directive Annex 1 this product is classified as a Monitoring and Control instrumentation product Do not dispose of electrical a...

Страница 9: ... 9 Cleaning 9 3 2 Connection 9 3 2 1 General Considerations 9 3 2 2 Electrical Considerations 9 3 3 Condensed Operating Instructions 10 3 4 Menu Structure 10 3 4 1 Home Screen 14 3 4 2 Setting a Resistance Value 14 3 4 3 Menu Key 14 3 4 4 Resistance Settings Menu 14 3 4 5 Standard Values Menu 15 3 4 5 1 Set Values 15 3 4 5 2 Standard Value 16 3 4 5 3 Decade Value 16 3 4 5 4 Percent Value 16 3 4 6 ...

Страница 10: ...tion 26 5 2 Capabilities 26 5 3 Address Switch and Communications Settings 26 5 4 GPIB Test Keyboard 26 Chapter 6 MAINTENANCE 28 6 1 Verification of Performance 28 6 1 1 Calibration Interval 28 6 1 2 General Considerations 28 6 1 3 Calibration Procedure 28 6 1 4 Adjustment of Internal Resistors 28 6 1 5 Calibration Menu and Process 30 6 1 6 Full Calibration adjustment 30 6 1 6 1 Automated versus M...

Страница 11: ...PRS 370 Series iv Table of Contents This page left intentionally blank ...

Страница 12: ...resistance source with excellent characteristics of stability temperature coefficient and power coefficient These features combined with a low virtually constant zero resistance make for very versatile instruments O p e r a t i o n i s b o t h l o c a l u s i n g convenient touch screen keypad and standard and IEEE 488 2 interfaces The PRS 370 employs very low resistance low thermal emf relays wit...

Страница 13: ...tector simulation It allows for higher precision and tight tolerances The PRS 370 is a precision resistance source with excellent characteristics of accuracy stability temperature coefficient and power coefficient It is direct reading without the need to subtract zero resistance The PRS 370 offers the option of true 4 wire Kelvin measurement for calibration applications and 2 wire mode for use in ...

Страница 14: ...e pre programmed into memory locations 1 to 4 Adjustment Automatic adjustment procedure utilizing a high pre cision DMM eliminates the requirement for manual trimming of resistors Terminals Front and rear connections each consisting of 4 low emf gold plated tellurium copper 5 way binding posts are used for HI and LO terminal pairs for CUR RENT and SENSE GND binding post is connected to the case to...

Страница 15: ...d label indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE Directive Annex I this product is classed as a Monitoring and Control instrumentation product Effects of Humidity Exposuretohumidity 60 foraperiodsofweeks can cause the resistors to drift some of that is reversible Itisrecomme...

Страница 16: ... or 6 digits Stability 50 ppm year Range 0 1 Ω 20 MΩ Thermal emf 15 μV Maximum load 2 A 200 V peak 0 5 W whichever applies first Resistors Precision wire wound and metal foil RTD tables PT 100 and PT 1000 tables for both Fahrenheit and Celsius are programmed into memory locations 4 RTD tables can be entered into memory to allow user selection of temperature and the correct value of resistance will...

Страница 17: ...SPECIFICATIONS PRS 370 Series 6 This page left intentionally blank ...

Страница 18: ...period of time it should be sealed in plastic and stored in a dry location It should not be subjected to temperature extremes beyond the specifica tions Extended exposure to such temperatures can result in an irreversible change in resistance and require recalibration 3 1 4 Bench and Rack Setup The PRS 370 should be placed on a stable flat work surface A front bail is provide to make viewing easie...

Страница 19: ...y replace fuse with the same type and rat ing UL CSA type T 0 8 A 250 V 5 x 20 mm 3 1 7 Environmental Do not operate the instrument in the presence of flammable gasses or fumes The PRS 370 is for indoor use only To ensure that the safety requirements the specifications and the measurement accuracy of the PRS 370 are met you must maintain the environmental temperature to within the speci fied range...

Страница 20: ...o 2 Wire See section 3 4 4 for more information The GND terminal on all models is connected to the case and to earth and chassis grounds This may be used as a shield terminal 3 2 2 Electrical Considerations The performance of the PRS 370 is directly affected by the quality of the connection to the system under test This is particularly true with the precision series models having higher accuracy a...

Страница 21: ... menu structure of the PRS 370 Menu items with a allow selection of a value or feature when ENTER pressed Menu items without have a submenu with more than 1 item Press MENU to display Main Menu and press the Back key to return to Home screen Below is the basic menu structure of the PRS 370 ...

Страница 22: ... Selection Front or Rear Standard Values RTD and Other Tables Settings Switch Mode Fast or Slow Resistance Settings Main Menu Set Value 0 1 ȍ Standard Value 1 Decade Value 0 1 ȍ Percent Value 0 001 Standard Values Resistance Settings Main Menu Probe Type None PT 100 C PT 100 F PT 1000 C PT 1000 F or Default User Selected RTD Table Settings Resistance Settings Main Menu Table Settings User Selected...

Страница 23: ...sistance Settings PRS DMM Settings Instrument Calibration only shown when Cal Enable pressed About PRS DMM Main Menu PRS DMM Off On Accuracy Slow Fast Instrument Settings Resistance Settings Main Menu DMM KS 3458A or FL8508A Address 1 30 Instrument Settings PRS 370 ...

Страница 24: ...nter Password Calibration History Instrument Calibration Main Menu Calibration History Calibrated Resistor R1 R43 Main Menu Resistance Settings PRS DMM Settings Instrument Calibration only shown when Cal Enable pressed About About Main Menu IET Labs Inc PRS 370 SN XX XXXXXXXX FW 1 15 HW 1 00 Reset to Default Yes or No Factory Reset Yes or No PRS 370 ...

Страница 25: ...temperature if RTD tables are used The main display also shows if the unit is configured for 4 wire mode or 2 wire mode in the upper left corner of the display after Resistance Figure 3 1 Home screen showing 2 Wire See section 3 2 1 for more information on 2 wire and 4 wire connections Pressing MENU will display the main menu Pressing INC or DEC will increment or decrement the resistance value bas...

Страница 26: ...hort modes When the resistance or temperature value is changed via the front panel or when a SOURCE DATA command is sent via the remote interfaces the intermediate value between previous value and the new value will either be a SHORT OPEN or undetermined in default switching case Standard Values and RTD and Other Tables Settings have separate menus which are described in the sections below 3 4 5 S...

Страница 27: ...es Association EIA and common values are E12 E24 E48 and E96 The number after the E denotes the number of resistance values the series contains per decade The series are logarithmic and are derived from the resistor tolerance resistors with a tighter tolerance can have more values in the series that won t overlap one another The series are sometimes referred to by the tolerance the two being relat...

Страница 28: ...e plus the resistance value multipled by the Percentage Value 3 4 6 RTD and Other Tables Menu Selecting the RTD and Other Tables menu allows use of tables There are 9 tables that can be selected The first 4 tables are defined as PT 100 C PT 100 F PT 1000 C and PT 1000 F where C is degrees Celsius and F is degrees Fahrenheit Press ENTER when Probe Type is highlighted to change between the various R...

Страница 29: ... name of the table can be entered If there are tables which need to be deleted then highlight the table and press DEL Once the table name is entered the table can be selected in User Functions and then a value of Units of Ω F C or K can be entered Use the key to move UP and the to move DOWN between User Functions Units and the Table The table shown in the bottom of Figure 3 7 can then be filled ou...

Страница 30: ...0 to control either a Fluke 8508A DMM or Keysight 3458A DMM and automatically adjust the output source resistance based upon measurements from the DMM 3 4 7 2 Accuracy Highlighting and pressing ENTER on Accuracy will allow selection of Slow or Fast accuracy in Figure 3 10 Figure 3 10 Accuracy When Slow is selected the DMM is configured for 7 5 digits and 6 5 digits in Fast This affects the speed o...

Страница 31: ...YES or NO to reset all parameters back to default The only item this reset does not affect is Calibration data 3 5 Thermal emf Considerations The PRS Series uses high quality low emf components Thermalemfisprimarilyattributable to the temperature difference between the leads of the relay and the contacts when temperature is applied to the coil This emf is of the order of 5 μV per relay but is not ...

Страница 32: ...y environments Always allow the instrument to stabilize at room temperature after unpacking or relocating the instrument Humidity should be maintained at laboratory conditions and typically 50 RH non condensing to minimize shifts in resistance 3 7 Remote Operation When the PRS is in REMOTE with front panel locked out an REMOTE indication will be shown in bottom half of the HOME screen Pressing the...

Страница 33: ...OPERATION PRS 370 Series 22 This page left intentionally blank ...

Страница 34: ... function in PRS DMM Mode 4 1 2 General Requirements PRS DMM Mode allows the PRS 370 to control a Fluke 8508Aor Keysight 3458ADMM via the IEEE interface A GPIB cable must be connected between the IEEE interface on the PRS 370 and the IEEE interface on the DMM The rear binding post of the PRS 370 should be connected to the front binding post of the DMM using a high quality 4 terminal kelvin connect...

Страница 35: ...tion of Slow or Fast accuracy in Figure 4 3 Figure 4 3 Accuracy When Slow is selected the DMM is configured for 7 5 digits and 6 5 digits in Fast This affects the speed or time required to perform adjustments in PRS DMM Mode Up to 5 iterations can be performed to adjust the source output in either accuracy 4 2 3 Configure Instrument Settings Highlighting and pressing ENTER on Instru ment Settings ...

Страница 36: ...reen Touching the resistance value brings up the entry screen for resistance or the numeric keypad can be used to enter the required resistance value as shown in Figure 4 7 Figure 4 7 PRS DMM Enter resistance value screen Enter the required resistance value and press one of the unit buttons To initiate an adjustment make sure IEEE and measurement cables are connected between the DMM and the PRS 37...

Страница 37: ...rtainties for TCAL 1 C Once fine Adjusting is complete the front connections are switched in and the rear con nections are switched off on the PRS 370 so both the DMM and DUT can be simultaneously connected to the PRS 370 When the source resistance is measured at the front terminals with a DMM 15 minutes lat er the source resistance shows no significant change This allows the PRS 370 to become a 4...

Страница 38: ...cols The standard is available on line at http www ieee org The SCPI standard provides a tree like series of standard commands for programmable instruments so that similar instruments by different manufacturers can be controlled by the same program SCPI information and a command reference are located in Appendix A Other tutorials are available on line consult IET for additional information A softw...

Страница 39: ...art up and it lets you enter just the data that you want to send to the device This program works with ICS Measurement Computing and National Instruments controllers To implement request a download of ICS_ GPIBkybd_Install zip from IET Labs Tech Support ...

Страница 40: ...er of hours at the nominal operating temperature of 23o C and at nominal laboratory conditions of humidity There should be no temperature gradients across the unit under test In the case of the PRS 370 a Kelvin type 4 wire test terminals should be used to obtain accurate low resistance readings Steps should be taken to minimize thermal emf effects by maintaining an even temperature and by using on...

Страница 41: ...performed to the PRS 370 Below is a list of the 43 resistors and their nominal resistance value These resistors should be within 1 of the nominal value for R1 to R5 and 0 1 of the nominal value for R6 and higher to ensure proper operation of the PRS 370 If the measured value of the resistor is outside of these specifications service is required The Calibration History will show each resistor value...

Страница 42: ...n Once this menu is displayed the CAL ENABLE key can be released Figure 6 2 Instrument Calibration Menu Full Calibration allows calibration of 1 or all 43 resistors Calibration of a single resistor is only available in manual mode Trace Calibration is only required if the PRS 370 requires service which involves changing of the relay board or internal wiring between binding post and the relay board...

Страница 43: ... Resistance Mode the process is manual If the PRS 370 is configured for PRS DMM Mode prior to going into the calibration routine for the measurement of all 43 resistors is done automatically with the PRS 370 controlling the DMM The PRS 370 should be configured as outlined in Chapter 4 This is a completely automated process If the PRS 370 is configured for source resistance mode the technician will...

Страница 44: ...his information can be used to look at the drift of each of the 43 internal resistors This data can be viewed in this menu If exporting this data is required see Appendix B remote commands which allow this data to be transferred to a PC Figure 6 6 Calibration History Screen From the Instrument Calibration menu select Calibration History and press ENTER The menu in Figure 6 6 will be shown Each tim...

Страница 45: ...370 Relay Board Relay Board Complete Not Shown PRS 370 CPU Board CPU Board Complete Not Shown PRS 370 Remote Interface USB IEEE Ethernet Interface Card 4 PRS 370 Touch Screen Touch Screen Not Shown PRS 370 BP Board Front Binding Post Board Front Not Shown PRS 370 BP Board Rear Binding Post Board Rear 5 PRS 370 FP Front Panel with Overlay and Bezel 6 PRS 370 Keypad Keypad and Keypad board 7 PRS 370...

Страница 46: ...here a reply is expected SCPI builds on the programming syntax of IEEE 488 2 to give the programmer the capability of handling a wide variety of instrument functions in a common manner This gives all instruments a common look and feel SCPI commands are not case sensitive The portion of commands shown in capitals denotes the abbreviated form of the keyword Either the abbreviated or whole keyword ma...

Страница 47: ...le register SRE Returns the read of the service request enable register ESE Set the event status enable register value SRE Set the service request enable register value RST Reset the buffer TST Self test query CLS Clear standard event status register OPC Operation complete OPC Operation complete query WAIT Wait until operation is complete before execut ing next command No Function in PRS PCS Power...

Страница 48: ...er is 0 to 9 RCL number Recalls resistance value from memory location 0 to 9 Number is 0 to 9 SYSTEM Commands SYSTem SERIal Ax xxxxxxxx Sets the serial number in non volatile memory serial num ber Serial number in the form Ax xxxxxxxx x 0 to 9 A Alpha can be A to Z SYSTem SERIal Returns Serial number SYSTem CONTrast number Sets system contrast specific number from 0 to 15 SYSTem KEY number Same as...

Страница 49: ... time the command CALI RES SET is used the resistor number is automatically incremented by 1 CALIbrate RESistance Returns resistance number to be calibrated Numbers 1 to 43 CALIbrate RESistance SET number Sets calibrated value of resistance in ohms into mem ory number is a floating point number representing the resistance value in ohms for the internal resistor being measured Each time the command...

Страница 50: ...display value currently shown in the PRS display SOURce RESistance display value Sets output resistance of the PRS 370 if table is set to None or the resistance value corresponding to the display value for the table selected Display value is in the form of number and decimal point representing desired display value This is the same as SOURce DATA command example SOURce RES 0 would set display valu...

Страница 51: ...OPEN SHORT CONFIGURE DMM Mode PRS 370 Only CONFigure PRSDMM ACCUracy SLOW FAST Sets resolution of DMM 7 or 6 respectively CONFigure PRSDMM ACCUracy Returns SLOW FAST CONFigure PRSDMM METEr 8508A 3458A Sets which DMM is used CONFigure PRSDMM METEr Returns string corresponding to selected DMM 8508A 3458A CONFigure PRSDMM ADDRess 1 30 corresponding to IEEE address 1 to 30 of DMM CONFigure PRSDMM ADDR...

Страница 52: ... number 0 1 10M CONFigure STDValue SET Returns number 0 1 10M CONFigure STDValue STD 1 5 10 EIA resistor CONFigure STDValue STD Returns 1 5 or 10 corresponding to EIA Resistor CONFigure STDValue DECade number 0 1 10M CONFigure STDValue DECade Returns number 0 1 10M CONFigure STDValue RADiometric number 0 001 10 where number is in CONFigure STDValue RADiometric Returns number 0 001 10 ...

Страница 53: ...ns the number 1 to 4 of currently selected table CONFigure TABLe ERASE Erases all rows in the table CONFigure TABLe ADD user_value resistance_value Adds a row to the end of table show ing user value and resistance value in ohms example CONF TABL ADD 20 95 means that if the table is selected and the operator enters 20 from the front panel the output will be set to 90 Ω CONFigure TABLe ADD Returns t...

Страница 54: ...PRS 370 Series APPENDIX B 43 This page left intentionally blank ...

Отзывы: