MAINTENANCE
PRS-370 Series
29
Chapter 6
MAINTENANCE
6.1 Verifi cation of Performance
6.1.1 Calibration Interval
The PRS Series instruments should be veri
fi
ed
for performance at a calibration interval of twelve
(12) months. This procedure may be carried out
by the user, if a calibration capability is available,
by IET Labs, or by a certified calibration
laboratory. If the user should choose to perform
this procedure, then the considerations below
should be observed.
6.1.2 General Considerations
It is important, whenever testing the PRS Series
Decade units, to be very aware of the capabilities
and limitations of the test instruments used. Such
instruments have to be signi
fi
cantly more accurate
than the speci
fi
ed accuracy for all applicable
ranges in order to perform this task, allowing for
a band of uncertainty of the instrument itself, the
test setup and the environment; consult IET Labs
for information.
It is important to allow both the testing instrument
and the PRS-370 to stabilize for a number of
hours at the nominal operating temperature
of 23
o
C, and at nominal laboratory conditions
of humidity. There should be no temperature
gradients across the unit under test.
In the case of the PRS-370 a Kelvin type 4-wire
test terminals should be used to obtain accurate
low-resistance readings.
Steps should be taken to minimize thermal-emf
effects, by maintaining an even temperature and
by using only low-emf connectors. Use of meters
with a “True Ohm” function is recommended.
Proper metrology practices should be followed
in performing this veri
fi
cation.
NOTE: The mode in the PRS-370 affects how
calibration is performed. If the PRS-370 is in
PRS-DMM Mode then the calibration process
is completely automated with the PRS-370
controlling the DMM during the calibration
process. If in Source Resistance Mode the
process is manual.
6.1.3 Calibration Procedure
Equipment Required:
Keysight 3458A, Fluke 8508A or equivalent
KK-100, TL-600 Kelvin Cable or equivalent.
Procedure:
Fluke 8508A or Keysight 3458A using 7 ½ digits
of resolution or better is connected to the binding
posts of the PRS-370 using a 4 terminal Kelvin
connection.
The gold shorting links must be removed on the
binding posts of the PRS-370.
Содержание PRS-370 SERIES
Страница 2: ...TEL 516 334 5959 800 899 8438 FAX 516 334 5988 www ietlabs com IET LABS INC ...
Страница 11: ...PRS 370 Series iv Table of Contents This page left intentionally blank ...
Страница 17: ...SPECIFICATIONS PRS 370 Series 6 This page left intentionally blank ...
Страница 33: ...OPERATION PRS 370 Series 22 This page left intentionally blank ...
Страница 54: ...PRS 370 Series APPENDIX B 43 This page left intentionally blank ...