adjustment. The calibration adjustment potenti-
ometer is located on the front panel beneath the
FUNCT
switch. Put the switches to the
F
,
Hi
,
A
, a
CAL 94dB
positions and adjust the poten-
tiometer for a display of exactly 94.0dB.
3.3.2 External Calibration
Attach an external sound level calibrator such
as the IET Labs 1562-A Sound Level Calibra-
tor (1kHz at 114 dB) or similar calibrator at
94 dB or 114 dB, completely over the micro-
phone. Set the meter to
F
,
Hi
, and
A
, and ad-
just the potentiometer for a display of exactly
94.0dB or 114 dB for 1562-A.
If the internal and external settings are not
within 1.5dB of each other, the meter will
require repair.