For convenience to the user, the pertinent 

specifi cations are given in an OPERATING 
GUIDE, shown typically in Figure 2.1, affi xed 
to the case of the instrument.

2.1 Specifi cations

Sound level range:

  35 to 130 dB re 20 µN/m2                  

Sound level accuracy:

 ±1.5 dB

Frequency range: 

 31.5 Hz to 8 kHz

Weighting:

  A and C

Output: 

 3.5 mm mini-jack

ac Output: >0.65 Vrms behind 600 Ω at full 
scale; will drive oscilloscope or low-imped-
ance headphones 
dc Output: 10 mV/dB

Display: 

3-1/2 digit LCD display, full scale 

sound-level range set by Range select switch, 
greater than 10 dB portion of scale for sound-
level comparisons on 2 ranges, and low end of 
sound-level range set by range select switch. 

Power:

 One 9V (6F22 or equivenent) alkaline 

battery supplied, provides 50+ hours of opera-
tion

Environmental:

  

 

Temperature: 

 0 to 50°C operating

-40 to +60°C storage, with batteries removed 

 

Humidity: 

<95% RH non-condensing

Weight:

 0.21 kg (7 oz) net, 1.4 kg (3 lb) ship-

ping

Dimensions:

 24 cm H x 6.8 cm W x 2.5 cm D 

(9.45” x 2” x 1”)

Accessories supplied:

 Instruction manual, 

windscreen, adjustment screwdriver and carry-
ing case

Additional Specifi cations
Standard:

  Conforms to ANSI S1.4-1983 Type 

2 and IEC61672-1:2013 Class 2

Dynamic range:

  55 dB with 2 ranges Lo and 

Hi

Ranges:

  Lo: 35 - 90 dB and Hi: 75 - 130 dB

Time weighting:

  rms response with fast 125 

mS and slow 1 s

Microphone:

  1/2” electret condenser micro-

phone

Resolution:

  0.1 dB

Calibration: 

 Internal oscillator at 1 kHz and 

94 dB

Recommended calibration interval: 

12 

months

Содержание 1565-E

Страница 1: ...REPRODUCTION AND DISTRIBUTION OF THIS TECHNICAL MANUAL IS AUTHORIZED FOR U S GOVERNMENT PURPOSES...

Страница 2: ......

Страница 3: ...charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the speci cations shall void this warranty IET shall not b...

Страница 4: ...E USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when working...

Страница 5: ...2 Connection 7 3 2 1 General Considerations 7 3 2 2 Electrical Considerations 7 3 2 3 Four Wire Kelvin Lead Connections 7 3 2 4 Thermal emf Considerations 7 3 3 Dial Setting 8 3 4 Environmental Condit...

Страница 6: ......

Страница 7: ...aped for comfort able hand held operation and tapered at the microphone end to minimize the effect of case diffraction It consists of the following elements a mi crophone to pickup sound an ampli er t...

Страница 8: ...range set by range select switch Power One 9V 6F22 or equivenent alkaline battery supplied provides 50 hours of opera tion Environmental Temperature 0 to 50 C operating 40 to 60 C storage with batteri...

Страница 9: ...FIGURE 2 1 Typical OPERATING GUIDE Af xed to Unit Please see label af xed to your unit...

Страница 10: ...Sound Level Meter Instruction Manual 2 Carry case for SLM and accessories 1565 E CASE 3 Windscreen 1565 E WS 4 Adjustment screwdriver 1565 0440 01 5 9 V battery 6F22 6 Phone plug for ac dc output 156...

Страница 11: ...s 3 2 2 FAST SLOW Response Time Select either FAST 125msec response or SLOW 1sec response measurement mode via the RESPONSE switch Selection of Fast or Slow is determined by the application and any di...

Страница 12: ...3 3 2 External Calibration Attach an external sound level calibrator such as the IET Labs 1562 A Sound Level Calibra tor 1kHz at 114 dB or similar calibrator at 94 dB or 114 dB completely over the mic...

Страница 13: ...1565 E worksheet Insert the microphone of the DUT fully into the transducer of the 1986 Set the DUT measurement speed to fast This speed will be used for all tests Turn on the DUT and select C weight...

Страница 14: ...uency Characteristics C Weighting 125 Hz 113 8 dB HI 112 3 115 3 dB 250 Hz 114 0 dB 112 5 115 5 dB 500 Hz 114 0 dB 112 5 115 5 dB 1 kHz 114 0 dB 112 5 115 5 dB 2 kHz 113 8 dB 112 3 115 3 dB 4 kHz 113...

Страница 15: ...window 1565 E 200 can be scratched Care should be taken to clean with soft cloth 4 3 3 Disassembly of the 1565 E The 1565 E can be disassembled by removing the two screws located on the rear of the 1...

Страница 16: ...565 E requires factory service 3 1565 E does not display 94 dB when set to internal calibration function of when using external calibrator Adjust calibration poten tiometer on front panel This is a no...

Страница 17: ......

Отзывы: