Error messages
43
DE
EN
13
Error messages
Blood pressure measurement errors
The blood pressure monitor indicates blood pressure measurement errors and communication errors by emitting
12 short acoustic signals.
Error
description
Possible cause
Remedy
Err 1
Arm moved during
measurement.
Keep your arm still during the measurement (Chapter 6.3).
Insufficient valid pulse rate
detected.
Put the blood pressure cuff on again (Chapter 6.2).
Err 2
Arm moved during
measurement.
Keep your arm still during the measurement (Chapter 6.3).
The blood pressure cuff is not
correctly positioned on your
arm.
Check the placement of the blood pressure cuff (Chapter
6.2).
Err 3
Blood pressure beyond the
measurement range.
If this notification appears continuously (or repeatedly), it is
possible that the blood pressure monitor is not suitable for
you. Contact your physician.
Strong arm movement.
Keep your arm still during the measurement (Chapter 6.3).
Err 4
IR communication error
If this error message is displayed continuously (or
repeatedly), contact your healthcare provider/physician or
the manufacturer.