Error messages
39
EN
Error
description
Cause
Remedy
Err 4
IR communication error
If this error message is displayed continuously (or repeatedly),
contact your healthcare provider or the manufacturer.
Err 5
Power pack or battery
voltage too low.
Change the batteries (
➔
Chapter 5.2
Battery contacts are
corroded.
Clean the battery contacts with a cotton cloth and a little alcohol.
Err 6
Air congestion
Check the blood pressure cuff for air congestion or a kink in the air
hose. If there is a kink in the air hose, unkink it, otherwise send the
blood pressure monitor to your healthcare provider or the
manufacturer.
Blood pressure cuff
incorrectly connected.
Connect the blood pressure cuff to the blood pressure monitor
(
➔
Chapter 6.2
Leak in the blood
pressure cuff or air hose
Replace the blood pressure cuff with the air hose.
Abr. (Abort)
Measurement cancelled
on request
Do not press this button when measuring, unless you want to abort
the measuring.
Содержание Tel-O-Graph BT plus
Страница 1: ...Operating manual EN Tel O Graph BT plus ...
Страница 51: ...Rev A 18 12 2018 ...