Battery Charge
The battery is charged with the charger supplied by IDMED. It can charge a flat battery in
less than 8 hours.
Icon for battery charge in progress or icon for
operation on the AC power adaptor
The battery can be charged regardless of its level. Whenever the ToFscan displays the flat
battery symbol (coloured red), the battery must be charged as soon as possible.
Charging is automatic, consequently when the charge is complete the ToFscan stops the
process.
Note :
Only qualified technicians or IDMED employees are authorized to carry out repairs or
maintenance operations on the battery.
Maintenance work on the battery is restricted to checking its charge cycle once every two
years. By doing this, it can be verified that the charge cycle does not exceed 8 hours (a change from
red to green for the battery charge gauge).
Battery replacement
It is recommended to change the battery every 2 years, during a periodic maintenance.