Service Manual ToFscan - Rev G
19
Tin all cables finishing by the ground cable. Make sure that all cable have the same length.
Slightly unwind the ground and put a heat-shrinkable sleeve.
The 6 leads have to be soldered to match the following pins configuration:
Pin 1
Pin 6
Pins configuration
Connector pins
Cable Leads
1
White
2
Brown
3
Green + shielding (Ground)
4
Yellow
5
Grey
6
Pink
Solder first the white cable.
Solder all cables counter-clockwise.
Test the proper functioning of the cable using a TOF-Controller before glueing all pieces together
Note:
If the results are not satisfying please check the soldering.
Apply a drop of glue on the cable collet, then insert the backnut inside the cable collet.
Screw the connector at the max and swipe out the excess of glue.
Apply a drop of silicon glue on the shell of the LEMO connector to keep the blue silicon protection in
place (ex.Silcoset 153).
Apply here
Let dry for 24 hours
After any servicing on ToFscan or its accessories, always validate its performance using the
TOF-Controller.
Cleaning
Caution:
Do not place the ToFscan or any of its parts or accessories inside an autoclave.
Under no circumstances must the ToFscan or any of its parts or accessories come into direct
contact with, be immersed in or filled with liquid.
The ToFscan and its parts and accessories are non-sterile devices. Under no circumstances must the
ToFscan or any of its accessories be sterilized.
The ToFscan must be cleaned and disinfected between each patient. Low-level disinfection is
generally sufficient.
The surfaces of the ToFscan and its accessories have to be cleaned with a lint-free cloth moistened