User Manual ToFscan
21
IV Servicing, Disinfection cleaning
Preventive servicing, Maintenance
In order to maintain its performance, it is strongly recommended that the following tests be carried
out on the device at least once every two years:
• Checking of the integrity of the casing, screen and labelling
• Checking of the battery charging process
• Checking of the condition of the electrode cable, its electrode clamp ends and its sensor clamp
• Checking for the value of current of electrical stimulations.
• Checking the sensor measures.
The lifetime of the ToFscan, under the required operating and maintenance conditions, is 5 years
(2 years for accessories).
Caution:
Only qualified technicians are authorised to carry out some repairs with the consent of IDMED.
Battery / Battery charge
Battery
The ToFscan includes a Lithium-Ion rechargeable battery. The battery is equipped with
thermal protection and short-circuits protection. At full charge, battery power lasts for approximately
1 month at a rate of 10 “TOF” stimulations per day (“Eco” mode activated).
Level and colour gauges indicates the level of charge of the battery
Battery level (green, yellow, red)
The battery is above 70% of its maximum capacity
The battery is between 20% and 70% of its maximum capacity
The battery is below 20% of its maximum capacity
The battery has a one-year warranty (its battery power at one year should be 50% more than its
estimated battery power). The normal service life of the battery is 2 years.
Note :
Only qualified technicians are authorised to carry out repairs or maintenance operations
with the consent of IDMED.
Battery Charge
The battery is charged with the charger supplied by IDMED. It can charge a flat battery in
less than 8 hours.
Icon for battery charge in progress or icon
for operation on the AC power adaptor
The battery can be charged regardless of its level. Whenever the ToFscan displays the flat
battery symbol (coloured red)
, the battery must be charged as soon as possible.
Charging is automatic, consequently when the charge is complete the ToFscan stops the
process.
Note :
Only qualified technicians or IDMED employees are authorized to carry out repairs or
maintenance operations on the battery.
Maintenance work on the battery is restricted to checking its charge cycle once every two
years. By doing this, it can be verified that the charge cycle does not exceed 8 hours (a change from
red to green for the battery charge gauge).