User Manual ToFscan
22
Cleaning and disinfection
Caution:
Do not place the ToFscan or any of its parts or accessories inside an autoclave.
Under no circumstances must the ToFscan or any of its parts or accessories come into direct
contact with, be immersed in or filled with liquid.
The ToFscan and its parts and accessories are non-sterile devices. Under no circumstances must the
ToFscan or any of its accessories be sterilized.
The ToFscan must be cleaned and disinfected between each patient. Low-level disinfection is
generally sufficient.
The surfaces of the ToFscan and its accessories have to be cleaned with a lint-free cloth moistened
with a Quaternary Ammonium Compound (QACs), isopropyl alcohol 70%. Before using any of these
solutions, refer to the manufacturer’s documentation and test on a reduced surface.
Example of recommended Quaternary Ammonium product:
- mikrozid
®
sensitive liquid from the manufacturer Schülke & Mayr GmbH.
Please check with your local authorized distributor or with the manufacturer which products are
available and approved in your country.
The ToFscan cable (electrode and/or sensor) must not come
into direct contact with, be
immersed in, splashed or filled with liquid and is to be cleaned in the same manner as the
ToFscan.
When cleaning the cables of the ToFscan, be careful not to create excessive traction on the splint that
could cause premature breakage of the wires inside the sheath.
Diagnostics / Possible dysfunctions
The table below summarises a list of possible dysfunctions and the solutions for resolving them.
Dysfunction
Solution
The device does not start or stops on its own
after a few seconds (Message “Low Battery”)
Set the device to charge (refer to the section
“Battery and battery charge”)
The sensor icon is displayed in grey
even
though the sensor is connected.
Check the condition of the cable and the
sensor. Disconnect the cable and reconnect to
the ToFscan
The ToFscan displays the impedance icon in red
(impedance too high)
Check the positioning of the electrodes and
how they couple with the patient (refer to
the section “Connection and positioning of
stimulation electrodes”).
Note:
Should the problem persists or is not solved by the actions listed in the table above, you
must contact the distributor. .
V End-of-life disposal / Recycling
In the interest of environmental safety, you are required to pass your used system on to a
collection body with the capability to treat devices containing electronic components and
Lithium-ion storage batteries.
To dispose of or recycle device components, contact a company specialising in the recycling of
electronic devices.