(e.g. ANIOS). Before using any of these solutions, refer to the manufacturer’s documentation
and carry out a test on a small area.
Example of recommended products :
- Incidin™ OxyWipe from the manufacturer Ecolab Inc.
Please check with your local authorized distributor or the manufacturer which products are
available and approved in your country.
The eyecup, the only part in contact with the patient, will be removed from the AlgiScan
and cleaned in the same way as the AlgiScan and then put back in place on the clean
AlgiScan. It can also be cleaned by autoclaving (maximum temperature 135°C). The eyecup
can withstand 50 autoclave washing cycles. Any alteration, degradation or modification of the
eyecup must lead to its replacement.
The eyecup must be cleaned or changed between each patient.
The AlgiScan must be cleaned and disinfected between each patient. Low-level disinfection
is usually sufficient.
The lens should always be free of stains or scratches to avoid the risk of distorting the
measurements made. It should be cleaned with a lint-free cloth and wiped carefully to avoid
any stains or reflections.
The electrode wire used with the AlgiScan should not come
into direct contact, be
submerged, sprayed or filled with a liquid and should be cleaned in the same way as the
AlgiScan.
When cleaning the cables of the AlgiScan, be careful not to create excessive traction on the
electrode cables that could cause premature breakage of the wires inside the sheath.
V Appendix 1
Troubleshooting
The table below is a list of possible malfunctions as well as the solution to be implemented
for their resolution.
Issue
Solution
Device will not turn ON, or turn OFF by itself
after a few seconds
Charge the battery by positioning the device
on the charger (See «Battery and battery
charging chapter»)
The LED charge indicator will not illuminate
while charging the device
Remove and reposition the device on its
charger.
The video image is blurred
Check the cleanliness
(see « Cleaning and Disinfection»)
The AlgiScan displays the red electrode
symbol and refuses to start recording in the
“Electrical Stimulation» mode
Check the positioning of the electrodes and
their coupling to the patient (refer to the
paragraph: “Connexion and positioning of
the stimulation electrodes”)