26
CAUTION
Chemical disinfection
•
Disinfection must be carried out in accordance with the general rules and
guidelines, taking into account explosion protection requirements.
•
Use a disinfectant that does not leave residues.
•
Use a disinfectant that is suitable for use on plastic.
•
Use a disinfectant that is compatible with the materials (air-guiding parts:
PC, silicone, TPU).
•
Use a disinfectant that is certified by the German Society for Hygiene and
Microbiology (Deutsche Gesellschaft für Hygiene und Mikrobiologie –
DGHM).
•
Alcohol and other solvents will cause the material to become matt or
brittle.
•
Never use corrosive polishes that contain solvents or abrasive
disinfectants.
•
Never expose the device to a disinfectant containing alcohol for more than
5 minutes.
•
Never immerse the base module or detachable module in disinfectant.
•
The use of a spray is not recommended because the disinfectant could
penetrate the mechanical parts of the Idiag P100.
If a disinfectant has been used which produces explosive gas mixtures, these
must have evaporated before you switch on the device.
Procedure
•
First clean all parts according to the instructions in Chapter 11.4 “Cleaning”
•
Disinfect the hand-held unit by wiping the surface with a lint-free cloth dampened with disinfectant.
•
Disinfect the air-guiding parts in a bath of disinfectant according to the instructions on the
disinfectant.
•
Open the breathing bag and rinse it with disinfectant.
•
After disinfection, rinse the immersed parts with water to remove disinfectant residues and ensure
that all parts are allowed to dry thoroughly.
Chemical
disinfection
Through immersion
Only by wiping the surface
Air guide
Mouthpiece
Breathing
bag
Base
module
Detachable
module
DANGER