Operator Manual
Aspect 100-4T and ISR
L100-2843 Rev. E
12
6.
EMERGENCY STOP PUSHBUTTON
➢
The red Emergency Stop Pushbutton is located in the center of the base
cover on the back of the table.
➢
Pressing this button until it is fully engaged will stop all motorized
movement by cutting off power from all system components.
➢
RESET: Restore the electrical functions by rotating the button a quarter turn.
7.
SETUP INSTRUCTIONS
7.1.
Setup
•
Hand operated controls for the movement of the tabletop are included with the table.
The operator should become familiar with the controls before using them.
•
External electrical power connection and disconnection is through the AC power cord
and outlet. The table will operate on 110 V~, 230 V~ or on internal battery backup
power (see section 8.2). When external electrical power is applied to the actuators
which control the movement of the tabletop, the green LEDs on the Handset Control
will illuminate.
•
The electrical power outlet used should be visible and accessible to the user. The
electrical power cord should be routed where it will not be subject to damage or be a
tripping hazard.
•
Check that the ground pin on the electrical cord plug is in good condition before each
time it is plugged in.
WARNING!
To avoid electric shock, plug the electrical power cord
into a properly grounded hospital grade outlet!
EMERGENCY
STOP
PUSHBUTTON
IMAGING END OF TABLETOP
PEDESTAL END OF TABLETOP