Operator Manual
Aspect 100-4T and ISR
L100-2843 Rev. E
11
5.
EMC (Electromagnetic Compatibility) STATEMENT
Portable and Mobile RF Communications Equipment can affect Medical Electrical
Equipment including this equipment. Use special precautions regarding EMC when these tables
are installed, operated, and maintained. EMC operating parameters for these tables are in the
SPECIFICATIONS section of this manual (Section 17).
The use of accessories, transducers and/or cables other than those specified, with the
exception of those sold by the manufacturer as replacement parts for internal components, may
result in increased emissions or decreased immunity of the equipment or system.
WARNING: This equipment should not be used adjacent to or stacked with other
medical electrical equipment and that if adjacent or stacked use is necessary, the
equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
WARNING: Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of this equipment, including cables connected to the
table. Otherwise, degradation of the performance of this equipment could result.
NOTE: The emissions characteristics of this equipment make it suitable
for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment might not offer
adequate protection to radio-frequency communication services. The user might need to take
mitigation measures, such as relocating or re-orienting the equipment.
If this equipment receives excessive electromagnetic interference, motion controls for
the table may be slow or unresponsive to user inputs. In this event:
1.
Verify the cause by turning nearby equipment off and retest motions. (Note: All
motions will be slowed when operating on battery backup as compared to full
AC power)
2.
If this problem is not resolved, immediately remove power to the equipment by
engaging the Emergency Stop Pushbutton as shown in section 6 of this manual.
3.
Notify IDI customer service using the contact information in section 18 of this
manual.
CAUTION
Applying more than 50 pounds of force to the side of
the tabletop can overcome the braking system causing
the tabletop to move.