Aspect Iso-Drive Q-CARD
CLEANING AND STORAGE:
1) Unplug power cord and
wrap it around the cord
holder.
2) Push in Emergency Stop
Pushbutton.
3) Set caster lock pedal to
LOCK position.
4) Clean and disinfect table.
WARNING! Read operator
manual prior to use
Scan to
Access
User
Manuals
Online
ACCESSORY RAILS
EMERGENCY
STOP BUTTON
CASTER LOCK
PEDAL
POWER CORD
HOLDER
CONNECTION PORT
FOR HAND CONTROL
(BOTH SIDES)
NON-SWIVELING
UNI-DIRECTIONAL
WHEELS
SWIVELING CASTERS
(CAN BE TOTAL LOCKED)
IMAGING END
OF TABLETOP
PEDESTAL END OF
TABLETOP
SETUP/PATIENT PREPARATION:
Moving table:
1) Unlock Casters using Caster Lock Pedal.
2) Position table to desired location.
3) Secure table mobility by locking Casters using Caster Lock
Pedal.
AC Power:
1) Plug power cord into properly grounded outlet.
2) Protect power cord from damage by casters or other
equipment.
Controls:
1) Emergency Stop Pushbutton disables all tabletop motion
control when engaged.
2) Ensure that hand control is connected to table.
3) If tabletop motion controls fail to operate, make sure that the
Emergency Stop Pushbutton is NOT pushed in.
4) Actuate controls to ascertain that power is present and table
movement functions properly.
EMERGENCY PROCEDURES:
Stop Tabletop motion:
•
Press Emergency Stop Pushbutton until it is fully engaged.
AC Failure:
•
Internal battery allows tabletop motions to function. Return
tabletop to proper position to unload patient.
CPR:
Return tabletop to level in both Trendelenburg and Iso-
Roll and lower table position to a comfortable height to
perform CPR.