SNAP

®

 Foal IgG Test

Test procedure

*

 *Before beginning any test, refer to the package insert for complete instructions.
 

†Anticoagulated whole blood (e.g., EDTA, heparin), serum, or plasma sample.

Test foals for immunoglobulin G (IgG) levels.

1.  Dilute sample.

 

Mix thoroughly 
and dispose of 
the first 

5–10 

drops

 from the 

bottle.

2.   Apply 

1 drop

 of sample 

over sample spot.

3.  Pour the 

entire bottle of 
conjugate into 
the sample well.

5.  Read the result at 

7 minutes

.

4. 

 

When color 

first

 appears 

in the activation circle, 
press firmly to activate. 
You will hear a distinct 
“snap.”

Activation  

circle

7

min

 

8°C

2°C

Sample spot

400 mg/dL 
calibrator spot

Sample spot

800 mg/dL
calibrator spot

Whole blood

serum, or 
plasma sample

For more information or to order this test, visit

 idexx.com/foal

.

400 mg/dL IgG 

Color intensity of sample spot is lighter than 

or equal to the 400 mg/dL calibrator spot.

400–800 mg/dL IgG 

Color intensity of sample spot is darker than 

the 400 mg/dL calibrator spot but lighter than 

or equal to the 800 mg/dL calibrator spot.

>800 mg/dL IgG 

Color intensity of sample spot is darker than 

the 800 mg/dL spot.

Содержание SNAP 4Dx Plus

Страница 1: ...SNAP Test Quick Reference Guide ...

Страница 2: ...its at a glance 3 SNAP Pro Analyzer 4 SNAP 4Dx Plus Test 5 SNAP Heartworm RT Test 6 SNAP Feline Triple Test 7 SNAP FIV FeLV Combo Test 8 SNAP cPL Test 9 SNAP fPL Test 10 SNAP Parvo Test 11 SNAP Giardia Test 12 SNAP Feline proBNP Test 13 SNAP Lepto Test 14 SNAP FeLV Antigen Test 15 SNAP Foal IgG Test 16 SNAP test cycling and shipping guidelines 17 ALL components MUST be at room temperature before t...

Страница 3: ...st Refrigerated until expiration 3 drops 4 10 SNAP FIV FeLV Combo Test 3 drops 4 10 SNAP cPL Test 3 drops 4 10 SNAP fPL Test 3 drops 4 10 SNAP Parvo Test Room temperature until expiration Coat entire swab tip with a thin layer of fecal material 5 8 SNAP Giardia Test Refrigerated until expiration Coat entire swab tip with a thin layer of fecal material 5 8 SNAP Feline proBNP Test 3 drops 5 10 SNAP ...

Страница 4: ... results screen and are saved on the analyzer automatically The color of the status light indicates the results of the run For more information or to order this product visit idexx com prohelp Icon Description Wireless Wi Fi signal status 5 gray bars best signal 5 red bars no signal Lost Wi Fi communication Connected and communicating with the IDEXX VetLab Station via a wired connection red icon i...

Страница 5: ...dispense 3 drops of sample and 4 drops of conjugate to the sample tube 2 Gently invert 3 5 times to mix 3 Pour the entire contents of the sample tube into the sample well of a SNAP 4Dx Plus device 5 Read the result at 8 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press fi...

Страница 6: ...device 5 Read the result at 8 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap 8 min Positive control High antigen indicator Negative control Whole blood serum or plasma sample For more information or to order this test vi...

Страница 7: ...Pour the entire contents of the sample tube into the sample well of a SNAP Feline Triple device 5 Read the result at 10 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap 10min Before beginning any test refer to the package ...

Страница 8: ... the sample tube into the sample well of a SNAP FIV FeLV Combo device 5 Read the result at 10 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap 10min Before beginning any test refer to the package insert for complete instru...

Страница 9: ... cPL device 5 Read the result at 10 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap 10min Before beginning any test refer to the package insert for complete instructions For more information or to order this test visit id...

Страница 10: ...mple well of a SNAP fPL device 5 Read the result at 10 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap 10min Before beginning any test refer to the package insert for complete instructions For more information or to order...

Страница 11: ...e result at 8 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap 8 min Test procedure Refer to package insert for complete storage requirements Before beginning any test refer to the package insert for complete instructions ...

Страница 12: ...nd release conjugate 2 Squeeze and release bulb 3 times to mix sample and conjugate 3 Squeeze bulb to dispense 5 drops into the sample well of a SNAP Giardia device 5 Read the result at 8 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a...

Страница 13: ...he sample well of a SNAP Feline proBNP device 5 Read the result at 10 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap 10min Before beginning any test refer to the package insert for complete instructions For more informat...

Страница 14: ...f the sample tube into the sample well of a SNAP Lepto device 5 Read the result at 10 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap Before beginning any test refer to the package insert for complete instructions 10min F...

Страница 15: ...at 10 minutes Activation circle 4 Use the SNAP Pro Analyzer to automatically activate the SNAP test and interpret the result OR When color first appears in the activation circle press firmly to activate You will hear a distinct snap 10min Test cats for feline leukemia virus FeLV 8 C 2 C Positive control FeLV Ag Negative control Whole blood serum or plasma sample For more information or to order th...

Страница 16: ... Read the result at 7 minutes 4 When color first appears in the activation circle press firmly to activate You will hear a distinct snap Activation circle 7 min 8 C 2 C Sample spot 400 mg dL calibrator spot Sample spot 800 mg dL calibrator spot Whole blood serum or plasma sample For more information or to order this test visit idexx com foal 400 mg dL IgG Color intensity of sample spot is lighter ...

Страница 17: ...NAP test and conjugate bottle listed below can be cycled up to 30 times For more information on each test visit idexx com snap If you have any questions contact IDEXX Customer Support at 1 800 248 2483 As part of our development process we perform stability studies including heat stress testing to ensure that our tests can withstand reasonable shipping stress In our heat stress testing process we ...

Страница 18: ...es Inc All rights reserved 06 0010825 00 All TM marks are owned by IDEXX Laboratories Inc or its affiliates in the United States and or other countries R E F E R E N C E LAB QUALITY I N T H E P A L M O F Y O U R H A N D SNAP ...

Отзывы: