Table 9. AC mA Test

 

Input Readings 

Step Current Frequency 

61-633 

61-635 

36.000mA 

50 Hz 

35.488 to 36.512 

35.632 to 36.368 

36.000mA 

400 Hz 

35.488 to 36.512 

35.632 to 36.368 

360.00mA 

50 Hz 

354.88 to 365.12 

356.32 to 363.68 

360.00mA 

400Hzz 

354.88 to 365.12 

356.32 to 363.68 

 
 

Testing the Amp

 

(A) Function 

To verify the accuracy in the ampere (A) measurement function, do the 
Following: 
1.   Connect the Calibrator to the 

A

 and 

COM 

inputs of the Meter. 

2.   Turn the rotary switch to A 
3.   Apply the inputs for steps 1-2 in Table 10. 
4.   For each input, compare the readings on the Meter display to the display readings in  
      Table 10. 
5.   If the display reading falls outside of the range shown in Table 10, the meter does not  
      meet specification. 
 

Table 10. DCA Test 

Input Readings 
Step Current 61-633 

61-635 

3.6000A 

3.5780 to 3.6220 

3.5888 to 3.6112 

9.000A 

8.9150 to 9.0850 

8.9420 to 9.0580 

 
6.   Press the blue button on the Meter to toggle to the AC amp measurement function. 
7.   Apply the inputs for steps 1-4 in Table 11. 
8.   For each input, compare the reading on the Meter display to the display readings in  
      Table 11. 
9.   If the display reading falls outside of the range shown in Table 11, the meter does not 
      meet specification. 

 
Table 11. ACA Test

 

Input Readings 

Step Current  Frequency 

61-633 

61-635 

3.6000A 

50 Hz 

3.5488 to 3.6512 

3.5632 to 3.6368 

3.6000A 

400 Hz 

3.5488 to 3.6512 

3.5632 to 3.6368 

9.000A 

50 Hz 

8.8120 to 9.1888 

8.8480 to 9.1520 

9.000A 

400 Hz 

8.8120 to 9.1888 

8.8480 to 9.1520 

 

Testing the Frequency Function 

To verify the accuracy of the Meter’s frequency function, do the following: 
1.   Connect the Calibrator to the 

V

Ω

 and 

COM

 inputs on the Meter. 

Note:  The accuracy of the Calibrator’s frequency function must be appropriate for the specified accuracy of the      
Meter. 

2.   Set the rotary switch to Hz. 
3.   Set the Function Generator for the square wave voltage and frequency for steps 1-5 of Table 12. 
4.   Compare the reading on the Meter display with the display reading shown in Table 12. 
5    If the display reading falls outside of the range shown in Table 12, the Meter does not  
      meet specification. 
 

 

Form Number TM61633-5 

Rev 5 Dec 2006 

Содержание 61-635

Страница 1: ...ollowing information Precautions and safety information Specifications Basic maintenance cleaning replacing the battery and fuses Performance test procedures Calibration and calibration adjustment pro...

Страница 2: ...t 8 Basic Maintenance 9 Opening the Meter Case 9 Replacing the Battery 9 Testing Fuses FS1 and FS2 9 Replacing Fuses 11 Cleaning 11 Input Terminals 11 Performance Tests 12 A Basic Operability Test 12...

Страница 3: ...nformation Use the Meter only as described in the Users Manual If you do not do so the protection provided by the Meter may be impaired Read the Safety Information page before servicing this product I...

Страница 4: ...proper Fuse To avoid fire hazard use only the fuse type and rating specified for this product Do not operate without covers To avoid personal injury do not apply any voltage or current to the product...

Страница 5: ...icated Automatic power off time Automatic backlit off 15 minutes Power source One 9V dry cell battery Maximum input voltage 1000V 750V AC CAT II Between V and COM Maximum floating voltage 1000V 750V A...

Страница 6: ...KHz N A 3 00 7d 20KHz 50KHz N A 5 00 8d 50KHz 100KHz N A 10 00 l0d 400V 40Hz 100Hz 1 20 5d 0 70 5d 100Hz 1KHz 2 00 5d 1 00 5d 1KHz 10KHz 3 00 6d 2 00 6d 10KHz 20KHz N A 3 00 7d 20KHz 50KHz N A 5 00 8d...

Страница 7: ...nput AC DC Current Same as AC RMS 1 00 8d C F Same as ACV PEAK HOLD 0 7 20 additional error for 10 of full scale RESISTANCE OHM 61 633 61 635 4000 4K 0 0 50 2d 0 30 2d 40KQ 400K0 0 50 2d 0 30 2d 4M0 0...

Страница 8: ...ACTOR Range 20 80 Resolution 0 1 Accuracy 6d 15Hz 10KHz 5Vp p Temperature Temperature 61 633 61 635 50 C 1200 C N A 1 C ld 100 C 50 C N A 2 C 1d 200 C 100 C N A 3 C 1d Physical Characteristics Charact...

Страница 9: ...egree 2 Do not operate in environments where conductive Pollutants may be present EC Declaration of Conformity Meets the intent of Directive 89 336 EEC for Electromagnetic Compatibility and Low Voltag...

Страница 10: ...tor AC Voltage Range 0 750V ac Accuracy 0 07 Basic Frequency Range 40 100KHz Accuracy 2 DC Voltage Range 0 1000V dc Accuracy 0 006 Basic Current Range 0 10A Accuracy AC 40Hz to 1KHz 0 08 Basic DC 0 02...

Страница 11: ...from the gasket 5 Hold the Meter display side up 6 Lifting up on the input terminal end disengage the case top from the gasket 7 Gently unsnap the case top at the display end Battery Replacement refe...

Страница 12: ...Page 10 Form Number TM61633 5 Rev 5 Dec 2006...

Страница 13: ...ter 9 Assemble the bottom case to top case using care not to damage battery wiring 10 Install the battery removed before and re assemble the battery cover Cleaning Warning To avoid electrical shock or...

Страница 14: ...Test To check the basic operability do the following 1 Turn the rotary switch to A and plug a test leads into A input terminal Unit display is of normal functional graphics if the fuse is good 2 Turn...

Страница 15: ...reading on the Meter display with the display reading shown in Table 2 If the display reading falls outside of the range shown in Table 2 the meter does not meet specification Table 2 AC Voltage Test...

Страница 16: ...frequency from step 1 2 in Table 4 7 Compare the reading on the Meter display with the display reading in Table 4 If the display reading falls outside of the range shown in Table 4 the Meter does not...

Страница 17: ...teps 1 2 Compare the Meter display readings to the display readings in Table 6 7 If the display reading falls outside of the reading shown in Table 6 the meter does not meet specification Table 6 LV I...

Страница 18: ...on the Meter 2 Turn the rotary switch to 3 Apply 3 000V The meter display should read approx 3 000V dc 4 Press Blue key to test function 5 Apply a 50 resistor to meter the built in beeper alarms Test...

Страница 19: ...7 Apply the inputs for steps 1 4 in Table 11 8 For each input compare the reading on the Meter display to the display readings in Table 11 9 If the display reading falls outside of the range shown in...

Страница 20: ...e range shown in Table 13 the Meter does not meet specification Table 13 Duty Factor Test Input Readings Step Duty Level 61 633 61 635 1 20 0 5Vp p 14 0 to 26 0 14 0 to 26 0 2 50 0 5Vp p 44 0 to 56 0...

Страница 21: ...and check rotary switch with correct position at same time Figure 3 4 Follow operation guides to calibrate one range 5 There are two status may show on button Case 1 push button and wait for process...

Страница 22: ...3 Description of CalWin300 window shown in Figure 4 1 Main Display Displays the main reading of the Meter 2 Connection Indicator Connection shows that the meter is connected to your Computer and the...

Страница 23: ...0 07 at least adjust VC3 to have a reading of 200 50 20 count 2 40V Range Set the calibrator to generate 20 000V 10KHz 0 07 at least adjust VC2 to have a reading of 20 050 20 count 3 4V Range Set the...

Отзывы: