8 - TROUBLESHOOTING

A number of commonly encountered issues are addressed in this section. It is recommended that you 

review these suggestions and follow directions to correct an issue before contacting customer service.

PROBLEM

SOLUTION

The screen is dark - nothing is showing.

Make sure the power cord is attached and plugged in.

Make sure the power switch on the back of the unit is turned 

on (press the “I” symbol on the switch).

Check and replace the fuses (see Section 6.3).

A warning screen is showing on the 

display.

Follow the directions and information provided on the 

screen before pressing the Clear button.

The red Safety Switch is glowing.

The button has been pressed. Press the button again - the 

light should turn off.

The volume on the unit is too loud or 

too soft.

Go to the Utilities screen and adjust the volume as desired.

A warning screen says no handpiece, 

or the wrong handpiece, is attached 

but I know the correct handpiece is 

connected.

Check the handpiece connector and make sure it is fully 

connected and secured.

Try attaching another handpiece and operating that 

handpiece. If the other handpiece works, then there may be 

a problem with the intended handpiece and should not be 

used.

The handpiece feels hot and/or 

the patient complains of a burning 

sensation.

Some warmth is normal as this indicates the procedure is 

working. If the heat is intolerable, use a lower power setting. 

Also ensure an ample amount of treatment oil is applied to 

the treatment area.

Tissues with less adipose tissue may get hotter relative to 

highly fatty tissue. Adjust the power intensity accordingly.

Treated tissue is red or appears 

swollen.

Slight redness is a natural outcome of the treatment. In case 

of extreme redness, reduce the power intensity.

There doesn’t appear to be any change 

in the treated tissue.

Increase the power intensity. Leaner tissue will not change as 

much since there is less fat to remove.

The unit is saying the temperature is 

too high.

Check and make sure the handpiece is not extraordinarily 

hot.

Make sure the unit is being used at ambient temperatures as 

described in Section 9.1.

WARNING MESSAGES

High Temperature - The unit and/or 

handpiece has reached excessive tem-

peratures.

Make sure the unit is being used in a place with ambient 

temperature (see Section 9.1).

Stop the treatment and allow the unit to sit for 5 minutes.

Timer Error - The unit’s internal clock is 

not working properly.

Turn the unit off, wait 10 seconds, then turn the unit back on. 

If this issue persists, contact Customer Service.

Current Fault - Excessive electrical 

power has occurred within the unit.

Turn the unit off, wait 10 seconds, then turn the unit back on. 

If this issue persists, contact Customer Service.

Содержание Ultimate Contour

Страница 1: ......

Страница 2: ...p 9 4 3 Registration 9 4 4 Handpiece Attachment 9 5 DEVICE OPERATION 10 5 1 Patient Preparation 10 5 2 Screens and Settings 10 5 3 Radio Frequency Treatment 13 5 4 Ultrasound Treatment 14 5 5 Concludi...

Страница 3: ...y Failure to read these instructions may result in damage to the device or risk of injury to you or to the patient It is advisable to keep this document with the device and refer to it often when ques...

Страница 4: ...s with a Body Mass Index BMI under 25 DO NOT use radio therapy on or near the face or neck DO NOT use on pregnant or nursing women on those who have given birth in the last 12 months or on those plann...

Страница 5: ...ls such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Ultimate Contour including cables specified by Ideal Curves Otherwise degradation of th...

Страница 6: ...onnected to an oxygen supply DO NOT use in the presence of flammable gases All treatments performed with this system must be subjected to the same clinical assessment as with traditional techniques Th...

Страница 7: ...e waist circumference of patients using the ultrasound portion This study evaluated a total of 42 individuals who were treated for 30 minutes at an intensity level of 4 during each treatment session T...

Страница 8: ...hock ONLY Federal law restricts this device to sale by or on the order of a physician CAUTION Make sure the unit is positioned so you can get easy access to the back panel of the unit in case you need...

Страница 9: ...e easily disconnected if needed If needed clean the unit and handpieces as described in Section 6 4 3 Registration Before beginning use of the Ultimate Contour it is highly recommended that you take a...

Страница 10: ...istent with the diameter of the handpiece This will help ensure that all of the treatment area is addressed and to help avoid overlapping treatment spots applying excessive energy to a section of tiss...

Страница 11: ...rently selected The relative power level of the energy The handpiece that is connected to the unit Details about specific differences will be discussed in Sections 5 3 To change the treatment time or...

Страница 12: ...stop the treatment early and then press the Standby button to return to the Standby screen the time will reset to the full amount of time you had selected CAUTION Do not lift up the handpiece during...

Страница 13: ...rth sweeping motion A good analogy to think of is mowing a lawn consistent sweeping patterns that just barely overlap slowly working from one side of the area to the other Apply slight pressure into t...

Страница 14: ...and the unit has stopped delivering energy lift the handpiece away from the patient tissue Set the handpiece aside Use clean towels to wipe away excess treatment oil from the skin Assess the treated a...

Страница 15: ...be treated with a disinfection solution such as a 1 3 glutaraldehyde solution or 0 2 peracetic acid solution The handpiece should be visibly wet and remain so for at least 5 minutes You may need to re...

Страница 16: ...th its accessories excluding the handpieces collectively defined as Product is warranted against defective materials and workmanship for a period of one 1 years from the date of purchase If the Produc...

Страница 17: ...you are having trouble with the Ultimate Contour system have questions regarding its use or need to order replacement parts feel free to contact a customer service representative at Ideal Curves 1 888...

Страница 18: ...ins of a burning sensation Some warmth is normal as this indicates the procedure is working If the heat is intolerable use a lower power setting Also ensure an ample amount of treatment oil is applied...

Страница 19: ...0 50 W cm2 1 00 W cm2 Level 3 0 75 W cm2 1 50 W cm2 Level 4 1 05 W cm2 2 00 W cm2 Ultrasound Focal Distance 6mm Ultrasound Focal Distance 6mm Ultrasound Focal Volume 2 12cm3 Ultrasound Entry Power 30...

Страница 20: ...to take mitigation measures such as relocating or re orienting the equipment Harmonic emissions IEC 61000 3 2 Not available Voltage Fluctuations Flicker emissions Not available The Ultimate Contour i...

Страница 21: ...in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey...

Страница 22: ...ration distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 d 3 5 V1 P d 3 5 E1 P GHz d 7 E1 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10...

Страница 23: ...23 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 24: ...Manufactured for Ideal Curves LLC by CAO Group 4628 West Skyhawk Dr West Jordan UT 84084 www ultimate contour com Rev 4 0 Mar 2021 1 888 861 7211...

Отзывы: