6.3 Fuse Replacement

The Ultimate Contour unit contains electrical fuses that can be replaced by the operator. 

FOLLOW THESE 

INSTRUCTIONS EXACTLY TO AVOID ELECTRICAL SHOCK.

 

1.  Unplug the power cord from the wall outlet, and detach the power cord from the back of the unit.

2.  Make sure the power switch is set to Off (press on the “O” symbol).

3.  Use a small blade screwdriver to dislodge the fuse compartment that is between the power switch and 

the power cord connection.

4.  Remove the fuse compartment. 

5.  Inside the compartment cover, remove the two fuses.

6.  Install the two new fuses.

 

WARNING: 

Make sure you install Type T2A250V fuses. Replacement fuses 

may be obtained from Ideal Curves (see Section 9.2). Failure to use the correct fuses may prevent the 

equipment from functioning properly, may damage the equipment, or may increase the risk of electric 

shock.

7.  Return the fuse compartment to the back of the unit. Make sure the 

wording on the compartment is right-side up. Be sure to press the com-

partment completely into place.

8.  Reconnect the power cord to the back of the unit. Plug the power cord into 

the wall outlet.

9.  Turn on the power switch (press the “I” symbol). If the fan turns on and the 

screen lights up, then the fuse replacement has been successful.

DANGER: 

Failure to follow these instructions may result in electric 

shock.

7 - LIMITED WARRANTY

7.1 Limited Warranty Statement

The Ultimate Contour device along with its accessories, excluding the handpieces (collectively defined 

as Product), is warranted against defective materials and workmanship for a period of one (1) years from 

the date of purchase. If the Product is leased from Ideal Curves or from any of its authorized distributors, 

consult your lease agreement for the terms and conditions of Product repair and/or replacement. Repair 

and/or replacement shall occur as directed by Ideal Curves or the authorized distributor. This warranty 

does not cover damage to the Product caused by accident, misuse or alteration/tampering of the Product. 

Warranty does not include labor or postage/freight charges. Warranty does not apply to the external finish 

of the Product. Ideal Curves reserves the right to make changes in design or to modify previously manufac-

tured Product. 

The handpieces are separately warranted against defective materials and workmanship for a period of one 

(1) year from the date of purchase. If the handpiece(s) is/are provided in association with a lease agreement 

with Ideal Curves or any of its authorized distributors, consult your lease agreement for the terms and con-

ditions of handpiece repair and/or replacement. 

7.2 Servicing

Replacement of the fuses is described in Section 6.3. No other portion or aspect of the Ultimate Contour 

system is serviceable by the end-user. Attempting to perform any other repair or service to the system 

may void the warranty. Ideal Curves will make available diagrams, parts lists, and schematics to authorized 

service personnel for servicing of this product. Although the Ultimate Contour is designed to provide con-

Содержание Ultimate Contour

Страница 1: ......

Страница 2: ...p 9 4 3 Registration 9 4 4 Handpiece Attachment 9 5 DEVICE OPERATION 10 5 1 Patient Preparation 10 5 2 Screens and Settings 10 5 3 Radio Frequency Treatment 13 5 4 Ultrasound Treatment 14 5 5 Concludi...

Страница 3: ...y Failure to read these instructions may result in damage to the device or risk of injury to you or to the patient It is advisable to keep this document with the device and refer to it often when ques...

Страница 4: ...s with a Body Mass Index BMI under 25 DO NOT use radio therapy on or near the face or neck DO NOT use on pregnant or nursing women on those who have given birth in the last 12 months or on those plann...

Страница 5: ...ls such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Ultimate Contour including cables specified by Ideal Curves Otherwise degradation of th...

Страница 6: ...onnected to an oxygen supply DO NOT use in the presence of flammable gases All treatments performed with this system must be subjected to the same clinical assessment as with traditional techniques Th...

Страница 7: ...e waist circumference of patients using the ultrasound portion This study evaluated a total of 42 individuals who were treated for 30 minutes at an intensity level of 4 during each treatment session T...

Страница 8: ...hock ONLY Federal law restricts this device to sale by or on the order of a physician CAUTION Make sure the unit is positioned so you can get easy access to the back panel of the unit in case you need...

Страница 9: ...e easily disconnected if needed If needed clean the unit and handpieces as described in Section 6 4 3 Registration Before beginning use of the Ultimate Contour it is highly recommended that you take a...

Страница 10: ...istent with the diameter of the handpiece This will help ensure that all of the treatment area is addressed and to help avoid overlapping treatment spots applying excessive energy to a section of tiss...

Страница 11: ...rently selected The relative power level of the energy The handpiece that is connected to the unit Details about specific differences will be discussed in Sections 5 3 To change the treatment time or...

Страница 12: ...stop the treatment early and then press the Standby button to return to the Standby screen the time will reset to the full amount of time you had selected CAUTION Do not lift up the handpiece during...

Страница 13: ...rth sweeping motion A good analogy to think of is mowing a lawn consistent sweeping patterns that just barely overlap slowly working from one side of the area to the other Apply slight pressure into t...

Страница 14: ...and the unit has stopped delivering energy lift the handpiece away from the patient tissue Set the handpiece aside Use clean towels to wipe away excess treatment oil from the skin Assess the treated a...

Страница 15: ...be treated with a disinfection solution such as a 1 3 glutaraldehyde solution or 0 2 peracetic acid solution The handpiece should be visibly wet and remain so for at least 5 minutes You may need to re...

Страница 16: ...th its accessories excluding the handpieces collectively defined as Product is warranted against defective materials and workmanship for a period of one 1 years from the date of purchase If the Produc...

Страница 17: ...you are having trouble with the Ultimate Contour system have questions regarding its use or need to order replacement parts feel free to contact a customer service representative at Ideal Curves 1 888...

Страница 18: ...ins of a burning sensation Some warmth is normal as this indicates the procedure is working If the heat is intolerable use a lower power setting Also ensure an ample amount of treatment oil is applied...

Страница 19: ...0 50 W cm2 1 00 W cm2 Level 3 0 75 W cm2 1 50 W cm2 Level 4 1 05 W cm2 2 00 W cm2 Ultrasound Focal Distance 6mm Ultrasound Focal Distance 6mm Ultrasound Focal Volume 2 12cm3 Ultrasound Entry Power 30...

Страница 20: ...to take mitigation measures such as relocating or re orienting the equipment Harmonic emissions IEC 61000 3 2 Not available Voltage Fluctuations Flicker emissions Not available The Ultimate Contour i...

Страница 21: ...in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey...

Страница 22: ...ration distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 d 3 5 V1 P d 3 5 E1 P GHz d 7 E1 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10...

Страница 23: ...23 THIS PAGE INTENTIONALLY LEFT BLANK...

Страница 24: ...Manufactured for Ideal Curves LLC by CAO Group 4628 West Skyhawk Dr West Jordan UT 84084 www ultimate contour com Rev 4 0 Mar 2021 1 888 861 7211...

Отзывы: