background image

16

6.3 Fuse Replacement

The Ultimate Contour unit contains electrical fuses that can be replaced by the operator. 

FOLLOW THESE 

INSTRUCTIONS EXACTLY TO AVOID ELECTRICAL SHOCK.

 

1.  Unplug the power cord from the wall outlet, and detach the power cord from the back of the unit.

2.  Make sure the power switch is set to Off (press on the “O” symbol).

3.  Use a small blade screwdriver to dislodge the fuse compartment that is between the power switch and 

the power cord connection.

4.  Remove the fuse compartment. 

5.  Inside the compartment cover, remove the two fuses.

6.  Install the two new fuses.

 

WARNING: 

Make sure you install Type T2A250V fuses. Replacement fuses 

may be obtained from Ideal Curves (see Section 9.2). Failure to use the correct fuses may prevent the 

equipment from functioning properly, may damage the equipment, or may increase the risk of electric 

shock.

7.  Return the fuse compartment to the back of the unit. Make sure the 

wording on the compartment is right-side up. Be sure to press the com-

partment completely into place.

8.  Reconnect the power cord to the back of the unit. Plug the power cord into 

the wall outlet.

9.  Turn on the power switch (press the “I” symbol). If the fan turns on and the 

screen lights up, then the fuse replacement has been successful.

DANGER: 

Failure to follow these instructions may result in electric 

shock.

7 - LIMITED WARRANTY

7.1 Limited Warranty Statement

The Ultimate Contour device along with its accessories, excluding the handpieces (collectively defined 

as Product), is warranted against defective materials and workmanship for a period of one (1) years from 

the date of purchase. If the Product is leased from Ideal Curves or from any of its authorized distributors, 

consult your lease agreement for the terms and conditions of Product repair and/or replacement. Repair 

and/or replacement shall occur as directed by Ideal Curves or the authorized distributor. This warranty 

does not cover damage to the Product caused by accident, misuse or alteration/tampering of the Product. 

Warranty does not include labor or postage/freight charges. Warranty does not apply to the external finish 

of the Product. Ideal Curves reserves the right to make changes in design or to modify previously manufac-

tured Product. 

The handpieces are separately warranted against defective materials and workmanship for a period of one 

(1) year from the date of purchase. If the handpiece(s) is/are provided in association with a lease agreement 

with Ideal Curves or any of its authorized distributors, consult your lease agreement for the terms and con-

ditions of handpiece repair and/or replacement. 

7.2 Servicing

Replacement of the fuses is described in Section 6.3. No other portion or aspect of the Ultimate Contour 

system is serviceable by the end-user. Attempting to perform any other repair or service to the system 

may void the warranty. Ideal Curves will make available diagrams, parts lists, and schematics to authorized 

service personnel for servicing of this product. Although the Ultimate Contour is designed to provide con-

Содержание 005-00036

Страница 1: ......

Страница 2: ...up 9 4 3 Registration 9 4 4 Handpiece Attachment 9 5 DEVICE OPERATION 10 5 1 Patient Preparation 10 5 2 Screens and Settings 10 5 3 Radio Frequency Treatment 13 5 4 Ultrasound Treatment 14 5 5 Concluding the Treatment 14 6 CLEANING AND MAINTENANCE 15 6 1 Handpiece Cleaning and Inspection 15 6 2 Unit Cleaning 15 6 3 Fuse Replacement 16 7 LIMITED WARRANTY 16 7 1 Limited Warranty Statement 16 7 2 Ser...

Страница 3: ...ly Failure to read these instructions may result in damage to the device or risk of injury to you or to the patient It is advisable to keep this document with the device and refer to it often when questions arise regarding its use and care If at any point you have any questions about the use or application of the Ultimate Contour feel free to contact a customer service representative at 1 888 861 ...

Страница 4: ...ns with a Body Mass Index BMI under 25 DO NOT use radio therapy on or near the face or neck DO NOT use on pregnant or nursing women on those who have given birth in the last 12 months or on those planning to become pregnant DO NOT use near plastic or metal parts within the body example implants stents ports Do not allow the transducers to contact hard surfaces when activated DO NOT use on patients...

Страница 5: ...als such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Ultimate Contour including cables specified by Ideal Curves Otherwise degradation of the performance of this equipment could result CAUTION Use of ultrasound on body parts other than the waist or on tissue that has a low adipose or fatty content may result in rapid heating tissue damag...

Страница 6: ...connected to an oxygen supply DO NOT use in the presence of flammable gases All treatments performed with this system must be subjected to the same clinical assessment as with traditional techniques The patient risk and medical history must be considered and understood before treatment Caution must be exercised regarding general medical conditions not already mentioned which might contraindicate a...

Страница 7: ...he waist circumference of patients using the ultrasound portion This study evaluated a total of 42 individuals who were treated for 30 minutes at an intensity level of 4 during each treatment session There were a total of 3 treatment sessions with 5 7 days interval between sessions The results show that on average a total waist reduction for the entire regimen was 2 0 inches The study included a 4...

Страница 8: ...Shock ONLY Federal law restricts this device to sale by or on the order of a physician CAUTION Make sure the unit is positioned so you can get easy access to the back panel of the unit in case you need to quickly turn the unit off or unplug it 4 INSTALLATION AND SET UP 4 1 Office Conditions The Ultimate Contour should be used in a controlled medical office environment The Ultimate Contour should b...

Страница 9: ...be easily disconnected if needed If needed clean the unit and handpieces as described in Section 6 4 3 Registration Before beginning use of the Ultimate Contour it is highly recommended that you take a moment to register at www ultimate contour com Registration facilitates customer service provides additional infor mation regarding the use of the Ultimate Contour managing patient expectations and ...

Страница 10: ...sistent with the diameter of the handpiece This will help ensure that all of the treatment area is addressed and to help avoid overlapping treatment spots applying excessive energy to a section of tissue which could cause thermal damage Make sure to use a marker that will not get wiped away by the treatment oils but will easily clean off after the treatment Use a tape measure and or skinfold calip...

Страница 11: ...rrently selected The relative power level of the energy The handpiece that is connected to the unit Details about specific differences will be discussed in Sections 5 3 To change the treatment time or the power press the Up arrow to increase the value or the Down arrow to decrease the value Minutes are selected in quantities of 10 minute incre ments from 10 to 40 minutes Intensity is selected at l...

Страница 12: ...u stop the treatment early and then press the Standby button to return to the Standby screen the time will reset to the full amount of time you had selected CAUTION Do not lift up the handpiece during treatment Stop the procedure first before lifting up the handpiece Do not touch the patient with your hands while the handpiece is active When 15 seconds remains in the treatment time the unit will s...

Страница 13: ...th sweeping motion A good analogy to think of is mowing a lawn consistent sweeping patterns that just barely overlap slowly working from one side of the area to the other Apply slight pressure into the skin so the handpiece has a sort of kneading effect on the tissue Most effective results is achieved when folds of skin press up in between the nodes on the handpiece surface Once the entire area ha...

Страница 14: ... and the unit has stopped delivering energy lift the handpiece away from the patient tissue Set the handpiece aside Use clean towels to wipe away excess treatment oil from the skin Assess the treated area There may be some redness or slight swelling of the treatment area Capture post treatment photos of the treated area using the same camera settings as were used for the pre treatment assess ment ...

Страница 15: ... be treated with a disinfection solution such as a 1 3 glutaraldehyde solution or 0 2 peracetic acid solution The handpiece should be visibly wet and remain so for at least 5 minutes You may need to reapply the disinfection solution to achieve this After the five minutes use a new dry towel or cloth to pat dry the handpiece Ensure the handpiece is fully dry before next use After each use look over...

Страница 16: ...ith its accessories excluding the handpieces collectively defined as Product is warranted against defective materials and workmanship for a period of one 1 years from the date of purchase If the Product is leased from Ideal Curves or from any of its authorized distributors consult your lease agreement for the terms and conditions of Product repair and or replacement Repair and or replacement shall...

Страница 17: ... you are having trouble with the Ultimate Contour system have questions regarding its use or need to order replacement parts feel free to contact a customer service representative at Ideal Curves 1 888 861 7211 or at the distributor from whom you obtained the system You may find it quick and helpful to review the Troubleshooting section of this manual Section 8 THIS SECTION INTENTIONALLY LEFT BLAN...

Страница 18: ...ains of a burning sensation Some warmth is normal as this indicates the procedure is working If the heat is intolerable use a lower power setting Also ensure an ample amount of treatment oil is applied to the treatment area Tissues with less adipose tissue may get hotter relative to highly fatty tissue Adjust the power intensity accordingly Treated tissue is red or appears swollen Slight redness i...

Страница 19: ... 0 50 W cm2 1 00 W cm2 Level 3 0 75 W cm2 1 50 W cm2 Level 4 1 05 W cm2 2 00 W cm2 Ultrasound Focal Distance 6mm Ultrasound Focal Distance 6mm Ultrasound Focal Volume 2 12cm3 Ultrasound Entry Power 30 2W Ultrasound Spatial Peak to Temporal Average Intensity 2 01W cm2 Ultrasound Pre Focal Temporal Average Intensity 1 32W cm2 Ultrasound Ratio Peak to Temporal Average Intensity 2 0 Ultrasound Peak Ra...

Страница 20: ...d to take mitigation measures such as relocating or re orienting the equipment Harmonic emissions IEC 61000 3 2 Not available Voltage Fluctuations Flicker emissions Not available The Ultimate Contour is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes Guidance an...

Страница 21: ... in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher freq...

Страница 22: ...aration distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 d 3 5 V1 P d 3 5 E1 P GHz d 7 E1 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 39 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicabl...

Страница 23: ...23 THIS PAGE INTENTIONALLY LEFT BLANK ...

Страница 24: ...Manufactured for Ideal Curves LLC by CAO Group 4628 West Skyhawk Dr West Jordan UT 84084 www ultimate contour com Rev 4 0 Mar 2021 1 888 861 7211 ...

Отзывы: