Setting reference 

To set reference press SET button bellow the reference you want to set. Current intensity 
value is stored as reference value.  

 

Using input filter 

Light detector has two input filters. Fast filter is intended for measurement of values 
changeable in time. It provides fast response with slightly increased noise level. Slow filter 
suppress noise but has rather slow response time. Thus it is suitable for measurement of in 
stable conditions when light intensity does not change fast. 

 

Setting input filter 

To adjust input filter press the 

FILTER

 button. The filter will toggle from slow to fast and 

vice versa with every pres of the button. 

 

Diagnostic modes 

The device provides four diagnostic modes which are useful for troubleshooting. The 
diagnostic modes are: zero offset, full scale, negative full scale, common mode rejection.  
 

Mode 

Description 

Zero offset 

Both differential inputs of A/D converter are switched to ground. 

Full scale 

Negative input of A/D converter is grounded. Positive is switched 
to the voltage reference. 

Negative full scale 

Positive input of A/D converter is grounded. Negative is switched 
to the voltage reference. 

Common mode rejection 

Both differential inputs of A/D converter are switched to voltage 
reference. 

 

Switching to diagnostic mode 

To switch to diagnostic mode press 

DIAG

 button. First diagnostic mode is selected. With 

every consecutive press of

 DIAG

 button device will switch to next diagnostic mode. After 

last diagnostic mode the device switch back to normal measurement mode. 
While in diagnostic mode you can switch to normal measurement mode anytime by pressing 

RANGE+

RANGE-

 or 

AUTO/MANUAL

 buttons. 

 

Basic maintenance tasks 

 

Replacing bulb in the light source 

To replace the bulb in the light source follow the steps described bellow. 

 

Switch off the device. 

 

Disconnect cable between control unit and the light source. 

 

Remove six screws on the back cover of the light source using allen key number 3. 

 

Loosen the cable gland so the cable can move freely through the gland. 

 

Remove the metal sheet cover and slide it away on the cable. 

 

Carefully remove the teflon back cover and slide it away on the cable. 

 

Remove internal light source assembly off the cooling jacket. 

 

Remove two screws which hold the bulb at position. Remove old bulb. 

Содержание FOTOMETER 2008 COOL

Страница 1: ...FOTOMETER 2008 COOL Light attenuation meter users guide ...

Страница 2: ...TIC MODE 6 SWITCHING BACK TO MANUAL MODE 6 OPERATING DEVICE IN MANUAL MODE 6 SWITCHING TO MANUAL MODE 6 SWITCHING BACK TO AUTOMATIC MODE 6 SWITCHING TO LESS SENSITIVE RANGE 6 SWITCHING TO MORE SENSITIVE RANGE 6 USING REFERENCES 6 SETTING REFERENCE 7 USING INPUT FILTER 7 SETTING INPUT FILTER 7 DIAGNOSTIC MODES 7 SWITCHING TO DIAGNOSTIC MODE 7 BASIC MAINTENANCE TASKS 7 REPLACING BULB IN THE LIGHT SO...

Страница 3: ...service must be provided by a qualified electrician only Device may be operated only in room environment between 15 and 40 C with non condensing humidity When moving device from cold environment to room temperature always allow the device temperature to accommodate to room environment to avoid moisture to condensate on the device before switching the device on Setting up the device Cable connectio...

Страница 4: ...nt air flow along the front window preventing the dust from burning chamber from sticking on the window Air purge inlet is located on the rear cover It is the topmost inlet on this side of unit The inlet must be connected to the compressed air source using appropriate pressure reduction valve and the regulation valve so the pressure and flow rate could be easily adjusted Maximum allowed pressure c...

Страница 5: ... range is already taken into account Range 100 Selected range 0 100 AU most sensitive Range 1000 Selected range 0 1000 AU Range 10000 Selected range 0 10000 AU Range Range 100000 Selected range 0 100000 AU less sensitive Reference 1 value in dB Relative change in intensity since the reference 1 was set expressed in dB Reference 2 value in dB Relative change in intensity since the reference 2 was s...

Страница 6: ...anual mode as well Operating device in manual mode In manual mode ranges must be switched manually by the user When range is manually switched device displays SWITCHING message and goes through the switching procedure in same manner as in automatic mode Switching to manual mode While in automatic mode you can press AUTO MANUAL button to switch to manual mode By pressing RANGE RANGE or DIAG while i...

Страница 7: ...e Negative full scale Positive input of A D converter is grounded Negative is switched to the voltage reference Common mode rejection Both differential inputs of A D converter are switched to voltage reference Switching to diagnostic mode To switch to diagnostic mode press DIAG button First diagnostic mode is selected With every consecutive press of DIAG button device will switch to next diagnosti...

Страница 8: ... printed circuit board on which the bulb is fixed Insert light source assembly back to cooling jacket Make sure it is properly positioned at the conical support flange at the front of cooling jacket The conical flange is visible through the front window Slide the teflon back cover on the cable to its place and press it firmly until the cover is aligned with the outer chassis Slide the metal sheet ...

Отзывы: