Operating device in automatic mode 

In automatic mode the device switch to more sensitive range, when intensity value decrease 
below 8 percent of current range. When intensity rise over 100 percent of current range, 
device automatically switches to less sensitive range. 
 

During switching, the device displays SWITCHING message on the display. During 

this period the intensity readout is not valid. On some conditions (slow filter selected, large 
sudden change in intensity) the intensity might not be settled when SWITCHING message 
disappear and additional waiting for valid value might be necessary. 

 

Switching to automatic mode 

While in manual mode, press 

AUTO/MANUAL

 button to switch to automatic mode. 

 

Switching back to manual mode 

While in automatic mode you can press 

AUTO/MANUAL

 button to switch back to manual 

mode. By pressing 

RANGE+

RANGE-

 or 

DIAG

 while in automatic mode you switch back 

to manual mode as well. 

 

Operating device in manual mode 

In manual mode, ranges must be switched manually by the user. When range is manually 
switched, device displays SWITCHING message and goes through the switching procedure in 
same manner as in automatic mode. 

 

Switching to manual mode 

While in automatic mode you can press 

AUTO/MANUAL

 button to switch to manual mode. 

By pressing 

RANGE+

RANGE-

 or 

DIAG

 while in automatic mode you switch to manual 

mode as well. 

 

Switching back to automatic mode 

While in manual mode, press 

AUTO/MANUAL

 button to switch to automatic mode. 

 

Switching to less sensitive range 

To switch to less sensitive range press 

RANGE-

 button. 

 

Switching to more sensitive range 

To switch to more sensitive range press 

RANGE+

 button. 

 

Using references 

To compute light attenuation directly during measurement there are two user references 
available. When reference is set it displays the ratio between reference value and current value 
in a logarithmic scale in dB. The value in logarithmic scale is computed using following 
formula: 
 





=

]

[

Re

]

[

log

10

]

[

AU

value

ference

AU

value

Intensity

dB

value

Log

 

Содержание FOTOMETER 2008 COOL

Страница 1: ...FOTOMETER 2008 COOL Light attenuation meter users guide ...

Страница 2: ...TIC MODE 6 SWITCHING BACK TO MANUAL MODE 6 OPERATING DEVICE IN MANUAL MODE 6 SWITCHING TO MANUAL MODE 6 SWITCHING BACK TO AUTOMATIC MODE 6 SWITCHING TO LESS SENSITIVE RANGE 6 SWITCHING TO MORE SENSITIVE RANGE 6 USING REFERENCES 6 SETTING REFERENCE 7 USING INPUT FILTER 7 SETTING INPUT FILTER 7 DIAGNOSTIC MODES 7 SWITCHING TO DIAGNOSTIC MODE 7 BASIC MAINTENANCE TASKS 7 REPLACING BULB IN THE LIGHT SO...

Страница 3: ...service must be provided by a qualified electrician only Device may be operated only in room environment between 15 and 40 C with non condensing humidity When moving device from cold environment to room temperature always allow the device temperature to accommodate to room environment to avoid moisture to condensate on the device before switching the device on Setting up the device Cable connectio...

Страница 4: ...nt air flow along the front window preventing the dust from burning chamber from sticking on the window Air purge inlet is located on the rear cover It is the topmost inlet on this side of unit The inlet must be connected to the compressed air source using appropriate pressure reduction valve and the regulation valve so the pressure and flow rate could be easily adjusted Maximum allowed pressure c...

Страница 5: ... range is already taken into account Range 100 Selected range 0 100 AU most sensitive Range 1000 Selected range 0 1000 AU Range 10000 Selected range 0 10000 AU Range Range 100000 Selected range 0 100000 AU less sensitive Reference 1 value in dB Relative change in intensity since the reference 1 was set expressed in dB Reference 2 value in dB Relative change in intensity since the reference 2 was s...

Страница 6: ...anual mode as well Operating device in manual mode In manual mode ranges must be switched manually by the user When range is manually switched device displays SWITCHING message and goes through the switching procedure in same manner as in automatic mode Switching to manual mode While in automatic mode you can press AUTO MANUAL button to switch to manual mode By pressing RANGE RANGE or DIAG while i...

Страница 7: ...e Negative full scale Positive input of A D converter is grounded Negative is switched to the voltage reference Common mode rejection Both differential inputs of A D converter are switched to voltage reference Switching to diagnostic mode To switch to diagnostic mode press DIAG button First diagnostic mode is selected With every consecutive press of DIAG button device will switch to next diagnosti...

Страница 8: ... printed circuit board on which the bulb is fixed Insert light source assembly back to cooling jacket Make sure it is properly positioned at the conical support flange at the front of cooling jacket The conical flange is visible through the front window Slide the teflon back cover on the cable to its place and press it firmly until the cover is aligned with the outer chassis Slide the metal sheet ...

Отзывы: