Device description 

Light attenuation meter FOTOMETER 2008 consists of modulated light source, lock-

in type light detector and control unit. Device measures intensity of light traveling from the 
light source to the light detector. It makes the device suitable for measurement changes of 
attenuation of light in the area between the light source and the light detector. Due to 
synchronization between light source modulation and the light detector the device is, to some 
extent, immune to ambient light and could be operated on normal day light conditions. 

The readout of intensity value is in linear arbitrary units (AU) and is not directly 

related to any engineering units. The device is intended only for measurement of relative 
changes in light intensity.  
 

Device has four decadic ranges and could be operated in two different modes. In 

manual mode

 the ranges must be switched manually by user. In 

automatic mode

 the device 

itself changes mode when intensity readout goes out of actual range. 
 

The light detector is equipped with two input filters. 

Fast filter

 provides quick 

response but slightly higher noise and is intended for measurement of unstable, changing light 
intensity. For measurement in stable, balanced conditions might be done using 

slow filter

 

which reduces noise but has very long settling time (up to one minute). 

 

Safety rules 

The control unit is powered by 230V AC which might be dangerous to life. Never remove the 
cover of control unit when the mains electricity is connected. Mains plug MUST be always 
removed before unscrewing the cover. Any service must be provided by a qualified electrician 
only. 
Device may be operated only in room environment between 15 and 40 C with non condensing 
humidity. When moving device from cold environment to room temperature always allow the 
device temperature to accommodate to room environment to avoid moisture to condensate on 
the device before switching the device on. 

 

Setting up the device 

 

Cable connections 

Mount control device to its place, preferably a 19 inch rack case. Connect cable with female 
connector between the control unit and the detector and cable with male connector between 
the control unit and the light source. Double check that cables are connected properly 
according to labels on control unit. Connect mains electricity and switch the device on. 

The light source and light detector must be mounted in such position that light source 

outlet is facing the light detector inlet directly. The light beam must go through the area where 
the light attenuation should be measured. Make sure that axes of light source outlet and light 
detector inlet are perfectly parallel. Using small piece of paper make sure that the light beam 
is aimed directly to the center of light detector inlet. 

 

Water cooling connections 

Detector and light source are equipped by water cooling which allows their usage in ambient 
temperatures as high as 150 °C. Both detector and light source have water inlet and water 
outlet on the rear cover of device. Inlet is situated on the bottom side, outlet on the top. The 
topmost tube is the air purge inlet (DO NOT connect water supply into the air purge inlet). 
 

Connect a silicone hose with 6mm internal diameter to the water inlets of both 

devices. Attach the other sides of hoses to the water supply. Water supply must be equipped 

Содержание FOTOMETER 2008 COOL

Страница 1: ...FOTOMETER 2008 COOL Light attenuation meter users guide ...

Страница 2: ...TIC MODE 6 SWITCHING BACK TO MANUAL MODE 6 OPERATING DEVICE IN MANUAL MODE 6 SWITCHING TO MANUAL MODE 6 SWITCHING BACK TO AUTOMATIC MODE 6 SWITCHING TO LESS SENSITIVE RANGE 6 SWITCHING TO MORE SENSITIVE RANGE 6 USING REFERENCES 6 SETTING REFERENCE 7 USING INPUT FILTER 7 SETTING INPUT FILTER 7 DIAGNOSTIC MODES 7 SWITCHING TO DIAGNOSTIC MODE 7 BASIC MAINTENANCE TASKS 7 REPLACING BULB IN THE LIGHT SO...

Страница 3: ...service must be provided by a qualified electrician only Device may be operated only in room environment between 15 and 40 C with non condensing humidity When moving device from cold environment to room temperature always allow the device temperature to accommodate to room environment to avoid moisture to condensate on the device before switching the device on Setting up the device Cable connectio...

Страница 4: ...nt air flow along the front window preventing the dust from burning chamber from sticking on the window Air purge inlet is located on the rear cover It is the topmost inlet on this side of unit The inlet must be connected to the compressed air source using appropriate pressure reduction valve and the regulation valve so the pressure and flow rate could be easily adjusted Maximum allowed pressure c...

Страница 5: ... range is already taken into account Range 100 Selected range 0 100 AU most sensitive Range 1000 Selected range 0 1000 AU Range 10000 Selected range 0 10000 AU Range Range 100000 Selected range 0 100000 AU less sensitive Reference 1 value in dB Relative change in intensity since the reference 1 was set expressed in dB Reference 2 value in dB Relative change in intensity since the reference 2 was s...

Страница 6: ...anual mode as well Operating device in manual mode In manual mode ranges must be switched manually by the user When range is manually switched device displays SWITCHING message and goes through the switching procedure in same manner as in automatic mode Switching to manual mode While in automatic mode you can press AUTO MANUAL button to switch to manual mode By pressing RANGE RANGE or DIAG while i...

Страница 7: ...e Negative full scale Positive input of A D converter is grounded Negative is switched to the voltage reference Common mode rejection Both differential inputs of A D converter are switched to voltage reference Switching to diagnostic mode To switch to diagnostic mode press DIAG button First diagnostic mode is selected With every consecutive press of DIAG button device will switch to next diagnosti...

Страница 8: ... printed circuit board on which the bulb is fixed Insert light source assembly back to cooling jacket Make sure it is properly positioned at the conical support flange at the front of cooling jacket The conical flange is visible through the front window Slide the teflon back cover on the cable to its place and press it firmly until the cover is aligned with the outer chassis Slide the metal sheet ...

Отзывы: