5 Putting Snap 41 into Operation

5.1  Connecting the Filling Tube

Screw in the filling tube by hand until some resistance against further turning 
can be felt. Tighten with your fingers only, do not use any tools.

Fig. 3: Connecting the filling tube

5.2  Mounting the Syringe Adapter

For some applications, filling with a syringe may be more convenient. To do 
so, mount the syringe adapter.

1. Remove the screw plug (see fig. 1).
2. Screw in the adapter Luer ¼" UNF by hand until some resistance against 

further turning can be felt. Tighten with your fingers only, do not use any 
tools.

5.3  Switching the Instrument On/Off

• To switch Snap 41 on, tap and hold the 

 key until the display lights up.

• To switch Snap 41 off, tap and hold the 

 key until the instrument is 

switched off.

TIP: 

For proper operation, cover the entire key area with your finger.

The instrument switches itself off after 5 minutes when idle and not moved.

Содержание Snap 41

Страница 1: ......

Страница 2: ...Instruction Manual and Safety Information Snap 41 Portable Alcohol Meter instrument software version from 1 0 Original Instructions...

Страница 3: ...rior notice All rights reserved including translation This document or any part of it may not be reproduced changed copied or distributed by means of electronic systems in any form print photocopy mic...

Страница 4: ...Setting the Language 20 7 2 Display Settings 20 7 3 Sound Settings 21 7 4 Enabling Disabling Gesture Control 21 8 Setting the Measuring Units 22 9 Checks and Adjustments 22 9 1 Performing a Water Chec...

Страница 5: ...11 8 System Information 33 12 Maintenance and Repair 34 12 1 Maintenance Performed by an Authorized Anton Paar Service Engineer 34 12 2 Repair Performed by an Authorized Anton Paar Representative 34 A...

Страница 6: ...our responsibility to establish health and safety practices and to determine the applicability of regulatory limitations Anton Paar GmbH only warrants the proper functioning of the instrument if no mo...

Страница 7: ...explosion The instrument is not explosion proof and therefore must not be operated in areas with risk of explosion General precautions Observe and adhere to your national safety regulations regarding...

Страница 8: ...e the legal require ments in your country 1 5 Conventions for Safety Messages The following conventions for safety messages are used in this document WARNING Description of risk Warning indicates a ha...

Страница 9: ...perform measurements of samples that are difficult to access A TFT display ensures clear visibility of results even in dark surroundings At the same time the backlight of the oscillator enables you to...

Страница 10: ...rument s temperature in order to keep the temperature influ ence on the measured result to a minimum The density and temperature values are the basis for the calculation of the alcohol content which i...

Страница 11: ...ement is not possible for liqueurs 3 Checking the Supplied Parts Snap 41 has been tested and packed carefully before shipment However damage may occur during transportation 1 Keep the packaging materi...

Страница 12: ...1 Rubber housing measuring cell 165214 1 Instruction manual English 185360 1 Filling tube 180 mm 172214 1 Adapter Luer UNF 172211 1 Syringes 2 mL Luer 10 pcs 58802 1 Allen key 3 mm DIN 911 58264 Tabl...

Страница 13: ...172246 Filling tube 600 mm 178843 Spare wristband for portable instruments 92477 Set carrying strap DMA 35 Snap 177301 Bluetooth USB adapter 99180 Carrying case DMA 35 Snap 41 177345 Alkaline battery...

Страница 14: ...Filling pump 4 Screw plug 5 Filling tube 6 Measuring cell 7 Power ON OFF key 8 LC display Power ON OFF key to switch the instrument on and off Arrow keys to navigate in the quick access area in menus...

Страница 15: ...14 E28IB004EN G Fig 2 Rear view of the instrument 1 Type plate with serial number 2 Screw of the battery compartment 2 1...

Страница 16: ...nient To do so mount the syringe adapter 1 Remove the screw plug see fig 1 2 Screw in the adapter Luer UNF by hand until some resistance against further turning can be felt Tighten with your fingers o...

Страница 17: ...The status bar shows the status of the instrument or a measurement If applicable a progress bar will show the progress of activities 1 Key functions 2 Status bar 3 Output field 4 Header 5 Battery char...

Страница 18: ...borts an operation Edit Switches into editing mode allowing you to change the current selection Live Unfreezes the last measuring result and returns to showing continuous live readings Menu Opens the...

Страница 19: ...keys to scroll through the content 6 5 Selecting from a List After you have switched into editing mode you see a selection list folded out 1 Use the arrow keys to select your choice 2 Tap OK to confir...

Страница 20: ...the position until the measurement starts Fig 5 Instrument turned to the right To abort a measurement 1 Turn the instrument to the left as shown in fig 6 and hold the position until a confirmation mes...

Страница 21: ...s well as the time after which the illumination of the display and the measuring cell will be automati cally dimmed To activate deactivate the automatic display rotation 1 Tap Menu and select Setup Di...

Страница 22: ...e instrument beeps whenever a key is tapped 1 Tap Menu and select Setup Sound 2 Tap Edit 3 Select On or Off from the selection list 4 Tap Back repeatedly to return to the main screen 7 4 Enabling Disa...

Страница 23: ...o verify that the instrument is measuring with adequate accuracy Perform a water adjustment if the water check advises you to do so IMPORTANT To obtain optimal results for the water check and the wate...

Страница 24: ...described in section 11 1 2 Rinse the measuring cell until no residues of cleaning agent are present anymore 3 Tap Menu and select Setup Water Adjustment 4 Fill in ultra pure water with 5 C to 30 C 4...

Страница 25: ...41 on your workbench TIP The pump lever with lock function supplied with the instrument is suited for any filling method However when you fill with the filling tube the pump lever without lock functio...

Страница 26: ...e this pump lever with the pump lever with lock function supplied with the instrument as described in section 11 6 1 Mount the syringe adapter see section 5 2 2 Lock the pump lever Push the pump lever...

Страница 27: ...r section 10 1 2 for filling with a syringe The instrument shows continuous live readings of measuring values 3 Start the measurement Activate Start function in the quick access area or use the corres...

Страница 28: ...ction 11 1 1 Cleaning Interval The cleaning interval strongly depends on the application For most samples a displacement of the previous sample by the next one will suffice In rare cases a cleaning af...

Страница 29: ...centration of the cleaning agent 11 1 4 Cleaning Agent Recommendations For cleaning the instrument use two cleaning liquids Cleaning liquid 1 dissolves and removes sample residues in the measuring cel...

Страница 30: ...n your application Dismounting the pump Fig 9 Dismounting the pump 1 Push the pump lever upwards with both thumbs see fig 9 2 Release the pump by a quarter turn counterclockwise 3 Pull out the pump Cl...

Страница 31: ...the pump lever down to fix the pump 11 3 Cleaning the Housing and the Display To clean the housing and the display use a soft cloth dipped in ethanol or warm water If necessary you can use a mild sol...

Страница 32: ...es Battery charge status A symbol in the header of the main screen indicates the battery charge status see fig 4 When the batteries are almost empty the icon starts to flash and after a short time the...

Страница 33: ...tteries for new ones 5 Reassemble all parts in reverse order 11 6 Exchanging the Pump Lever 1 Dismount the pump as described in section 11 2 2 The pump lever sits on the two pivot pins Pull it off car...

Страница 34: ...d select Send it to Bluetooth 3 On Snap 41 tap Menu and select Setup Software Update 4 Tap Start to start the automatic update procedure 11 8 System Information The instrument holds system information...

Страница 35: ...trument are generally excluded from the warranty as well 12 2 Repair Performed by an Authorized Anton Paar Representative In case your instrument needs repair contact your local Anton Paar repre senta...

Страница 36: ...ion 0 01 v v Temperature Measuring range 5 C to 30 C 41 F to 86 F Accuracy 0 2 C 0 4 F Repeatability s d b 0 1 C 0 2 F Resolution 0 1 C 0 1 F Sample volume 2 mL Sample temperature max 100 C 212 F Ambi...

Страница 37: ...aterial Part Borosilicate glass measuring cell filling pump Kalrez flat seal between connection block and measuring cell PP polypropylene housing PTFE polytetrafluoroethylene filling tube PVDF polyvin...

Страница 38: ...E28IB004EN G 37 Appendix B Declarations of Conformity...

Страница 39: ...38 E28IB004EN G...

Страница 40: ...Appendix C Menu Tree Setup Water Adjustment Factory Adjustment Display Settings Gesture Control Sound Language Software Update System Information...

Отзывы: