E28IB004EN-G
23
6. After the procedure has finished, the check result is displayed.
If the deviation from the target value exceeds defined limits, you will see
a recommendation to perform a water adjustment.
Tap <OK>.
7. Tap <Back> repeatedly to return to the main screen.
9.2 Performing a Water Adjustment
Use freshly degassed ultra-pure water.
IMPORTANT:
To obtain optimal measuring results, perform the water
adjustment at a temperature close to the temperature during routine
measurements.
1. Clean the measuring cell as described in section 11.1.
2. Rinse the measuring cell until no residues of cleaning agent are present
anymore.
3. Tap <Menu> and select
Setup > Water Adjustment
.
4. Fill in ultra-pure water with 5 °C to 30 °C (41 °F to 86 °F).
5. Tap <Start>.
6. After the procedure has finished, the density deviation from the previous
adjustment is displayed.
7. To save the new adjustment, tap <Save>.
To refuse the adjustment, tap <Cancel>.
9.3 Reset to Factory Adjustment
If you want to undo all your adjustments, reset the instrument to factory
adjustment.
1. Tap <Menu> and select
Setup > Factory Adjustment
.
2. Tap <Yes> to reset the instrument to factory adjustment.
The factory adjustment will be restored.
3. Tap <Back> repeatedly to return to the main screen.