7 Performing a Measurement

E15IB001EN-N

17

7.2.1  Filling with a Syringe (Luer Tip)

IMPORTANT: 

Do not use syringes that contain 

lubricants. The lubricants can dissolve into your 
sample and lead to a systematic measuring error.

Filling position

• To achieve best measurement results, fill with 

the syringe attached to the syringe holder.
This is the standard filling position.

Only in the case that the available amount of 
sample is very limited:

- Remove the adapter Luer 1/4" UNF from the 

syringe holder (1, fig. 2) and screw it into the 
sample inlet (4, fig. 2) of the instrument 
(instead of hose 3, fig. 2).

- Use a 2 mL syringe for filling.

To fill with a syringe (Luer tip)

1. Connect the syringe (filled with sample) to the 

adapter Luer 1/4" UNF (1, fig. 2) on the syringe 
holder (or to the adapter mounted directly at the 
sample inlet of the instrument).

2. Push the plunger of the syringe slowly and 

steadily until a drop emerges from the sample 
outlet adapter.

3. Leave the syringe in the filling position during 

the measurement.

7.2.2  Filling with Optional Peristaltic Pump

See the manual of Xsample 200 High-Resistance.

7.3  Measurement Procedure

IMPORTANT: 

If the adjustment temperature differs 

from the measuring temperature, give the measur-
ing cell sufficient time for conditioning (approx. 
1 hour per 10 °C/18 °F difference).

See section 8.2 for recommendations when to 
perform an adjustment before a measurement.

Particularly after temperature jumps the measure-
ment results will be slightly drifting before they 
become stable. This is due to the intrinsic properties 
of glass, which the measuring cell is made of.

If you want to achieve best measurement results, 
give the measuring cell and sample enough time for 
temperature conditioning.

• As a rule of thumb allow for a conditioning time 

of 1 minute per every 10 °C (18 °F) difference 
between the temperature of the filled sample 
and the set measuring temperature.

• Additionally set a hold time (see section 6.3.4) of 

4 min (DMA 501) / 5 min (DMA 1001) if you do 
not check density stability in your measurement.

If you measure density that is not viscosity-correct-
ed, allow for a longer conditioning time: approx. 
5 min if you check density stability, or approx. 10 min 
if you don’t.

Initiating the measurement

• When you have verified that all measurement 

settings are correct, and the sample has been 
filled, tap 

MEASURE

.

The measurement will be performed according 
to your settings. The progress bar shows the 
progress of the measurement.
When the measurement is finished, the results 
are displayed.
The results together with the camera image and 
a time stamp are automatically saved in the 
data memory.

• Tap

 

WARNING

Risk of injuries by dangerous liquids leaking 
or spurting out

If you push the plunger of the syringe too hard, 
or if, e.g., the measuring cell breaks, dangerous 
liquids may leak or spurt out of ruptures causing 
injuries.
• Wear safety goggles and protective gloves 

when you fill dangerous liquids.

 

WARNING

Risk of fire and poisonous substances

In the very rare case of a malfunction of the 
temperature control, the measuring cell could 
heat up to 150 °C (302 °F).
• Set the safety temperature limiter (9A, fig. 1) on 

the rear of the instrument to a temperature that 
is safe for your sample (use a screwdriver).

Содержание Anton Paar DMA 501

Страница 1: ...Instruction Manual and Safety Information DMA 501 DMA 1001 Density Meter instrument software version from 4 1 0 Original Instructions...

Страница 2: ...and its contents may be changed or amended by Anton Paar at any time without prior notice All rights reserved including translation This document or any part of it may not be reproduced changed copie...

Страница 3: ...13 6 Measurement Settings Products 14 6 1 Products 14 6 2 Measurement Settings 14 6 2 1 Dashboards 14 6 2 2 Reports 14 6 3 Measurement Parameters 14 6 3 1 Set Temperature 14 6 3 2 Measurement Perform...

Страница 4: ...ed Anton Paar Service Engineer 25 10 2 Repair Performed by an Authorized Anton Paar Representative 25 Appendix A Technical Data 26 A 1 Specifications 26 A 2 Instrument Data and Operating Conditions 27...

Страница 5: ...s are taken based on the results 1 2 Installation and Use The installation procedure shall be carried out only by authorized persons who are familiar with the installation instructions Use only access...

Страница 6: ...e of flammable gases and vapors Provide fire extinguishing equipment Transportation Empty the measuring cell and all hoses before you move or lift the instrument To carry the instrument put one hand i...

Страница 7: ...era image U View Intuitive user interface Operate the instrument via the touchscreen user interface giving you the look and feel of a smart phone If you prefer you can optionally connect an external k...

Страница 8: ...er jack 8 On Off switch 9 Safety temperature limiter A and reset button B 10 Type plate with serial number P N mat no Air connectors on the rear 11 Air pump 4 5 mm barbed inlet C outlet D 12 Drying of...

Страница 9: ...nough room approx 10 cm 4 in between the cooling fins on the rear of the instrument and any wall Also ensure that the heat can freely dissipate Do not install the instrument in a cabinet The instrumen...

Страница 10: ...f optional parts refer to the particular instructions coming with the part or to the corresponding section in the reference guide 3 2 6 Power Connection 1 Connect the power supply to the DC power jack...

Страница 11: ...but during transport the density adjustment may have been compromised If both checks succeed your instrument is ready for routine measurements Else perform an air water adjustment as described in sec...

Страница 12: ...Use time server to Yes The Time servers input field replaces the input fields Date and Time c The Time servers setting depends on your network settings If you have set up an automatic network configur...

Страница 13: ...details You can place the following widgets onto a dash board of the instrument the live camera image of the measuring cell U View measurement information giving information about the current status...

Страница 14: ...Yes is the standard setting Select No if the detection of filling errors by FillingCheck obviously contradicts what you see in the live camera image TIP FillingCheck provides excellent support to the...

Страница 15: ...in the menu then select the specific product b Check the measurement settings or edit them then tap Create Measurement To define all measurement settings on the fly and perform a measurement without...

Страница 16: ...ll corrode the wetted parts Sample leaking from corroded parts may cause serious injuries Before you fill any sample or cleaning liquid in particular hazardous or flammable chemicals into the instrume...

Страница 17: ...is due to the intrinsic properties of glass which the measuring cell is made of If you want to achieve best measurement results give the measuring cell and sample enough time for temperature condition...

Страница 18: ...large enough to leave some space for creating the vacuum Only up to two thirds of the syringe should be filled up Boiling the sample 1 Boil the liquid for several minutes to remove dissolved gas 2 Fi...

Страница 19: ...lid again Examine the camera image included in the results to check that the water has been filled bubble free Check the quality of the water Clean the measuring cell thoroughly If above actions do no...

Страница 20: ...TIP The atmospheric pressure that you get from a local weather station is usually not the absolute atmospheric pressure but a calculated atmospheric pressure value referenced to sea level b Enter the...

Страница 21: ...nts See the manual of Xsample 200 High Resistance for a description of the cleaning and drying proce dure with the optional peristaltic pump 1 Select Cleaning in the menu 2 Rinse the measuring cell wi...

Страница 22: ...glass surface which are difficult to remove If you store the instrument for less than one day the measuring cell can be filled with ultra pure water WARNING Before using any cleaning agents for the in...

Страница 23: ...he instrument are generally excluded from the warranty as well 10 2 Repair Performed by an Autho rized Anton Paar Representative In case your instrument needs repair contact your local Anton Paar repr...

Страница 24: ...ideal conditions and for low densities viscosities 0 001 g cm3 0 0001 g cm3 Temperature Measuring range 15 C to 40 C 59 F to 104 F 15 C to 60 C 59 F to 140 F Repeatability s d a 0 1 C 0 2 F 0 02 C 0 0...

Страница 25: ...degree 2 Overvoltage category II for the supplied AC adapter Display 7 TFT WVGA 800x480 px PCAP touchscreen Controls touchscreen optional keyboard mouse and barcode reader Interfaces 1x Ethernet 3x US...

Страница 26: ...Appendix B EU Declaration of Conformity Appendix B EU Declaration of Conformity...

Отзывы: