1.8.2 Analog Output
The Analog Output calibration procedure is as follows:
1.
Warm up the module for 30 minutes.
2.
Set the Type Code to the type you wish to calibrate. Refer to
Section 2.23 for details.
3.
Enable calibration. Refer to Section 2.47 for details.
4.
Set the zero Analog Output voltage/current. Refer to Section 2.5
for details.
5.
Check the meter and trim the output until zero output is achieved.
Refer to Section 2.11 for details.
6.
Send the Analog Output Zero Calibration command. Refer to
Section 2.7 for details.
7.
Set the span Analog Output voltage/current. Refer to Section 2.5
for details.
8.
Check the meter and trim the output until the span output is
achieved. Refer to Section 2.11 for details.
9.
Send the Analog Output Span Calibration command. Refer to
Section 2.9 for details.
Notes:
1.
For Analog Output channels, calibration must be performed for
each channel individually, so the calibration voltage/current
source should be connected to the specific channel that is to be
calibrated.
2.
When calibrating Types Codes 0 and 1, the jumper for the
corresponding channel
should be set to the “current input”
position.
3.
Calibration voltages and currents are shown below.
Calibration Voltages/Current:
Type Code
0
1
2
3
4
5
Zero Output
0 mA
4 mA
0 V
0 V
0 V
0 V
Span Output
+20 mA
+20 mA
+10 V
+10 V
+5 V
+5 V
Содержание M-7026
Страница 8: ...M 7026 User Manual Rev 1 5 Date 2016 03 10 8 1 3 Pin Assignments...
Страница 12: ...M 7026 User Manual Rev 1 5 Date 2016 03 10 12 1 5 Block Diagram...
Страница 13: ...M 7026 User Manual Rev 1 5 Date 2016 03 10 13 1 6 Application Wiring...
Страница 39: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 39 Related Topics Section 1 8 1 Analog Input Calibration...
Страница 41: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 41 Related Topics Section 1 8 2 Analog Output Calibration...
Страница 43: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 43 Section 1 8 1 Analog Input Calibration...
Страница 45: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 45 Related Topics Section 1 8 2 Analog Output Calibration...
Страница 47: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 47 Section 5 1 INIT Mode...
Страница 57: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 57 Section 1 9 5 Analog Output Type Codes and Data Format...
Страница 65: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 65 Section 1 9 5 Analog Output Type Codes and Data Format...
Страница 90: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 90 Section 5 2 Dual Watchdog Operation...
Страница 106: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 106 Section 2 35 AAS1 Related Topics Section 1 8 Calibration...
Страница 113: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 113 Section 2 68 AARECi...
Страница 118: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 118 Section 2 54 AACH Section 2 69 AARH Section 2 71 AARHi...
Страница 123: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 123 Section 2 57 AACL Section 2 72 AARL Section 2 74 AARLi...
Страница 127: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 127 Section 2 62 AADODD...
Страница 129: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 129 Related Commands Section 2 61 AADI...
Страница 137: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 137 AARAO Section 2 70 AARHCi Section 2 73 AARLCi...
Страница 141: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 141 Related Commands Section 2 53 AACECi...
Страница 143: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 143 Section 1 9 4 Analog Input Type Codes and Data Format...
Страница 180: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 180...
Страница 183: ...M 7026 M 7026 User Manual Rev 1 5 Date 2016 03 10 183...