T-CAM 80 P-SERIES/T-CAM 160 XT USER MANUAL
23
9-17 Temperature Compensation
The imager has temperature compensation to correct surrounding environmental sources
of energy. To get a more accurate temperature measurement, set the temperature
compensation.
Set the temperature compensation as follows:
1.
Press the Menu button.
2. Highlight Emissivity from the Menu.
3. Press the Select button. A sub-menu will open.
4. Highlight Temp. Comp. in the sub-menu.
5. Press the Select button.
6. Press the Up or Down button to modify the temperature compensation.
7. Press the Select button to confirm. Press Close to exit without changes.
8. Press Close to exit the menu.
9-18 Alarm
The Alarm enables threshold settings for temperature screening. An audible will sound and
the Level & Span bar will flash red on-screen if the threshold is exceeded.
Set the threshold as follows:
1.
Press the Menu button.
2. Highlight Emissivity from the Menu.
3. Press the Select button. A sub-menu will open.
4. Highlight Alarm in the sub-menu.
5. Press the Select button.
6. Press the Up or Down button to modify the Alarm settings.
7. Press the Select button to confirm. Press Close to exit without changes.
8. Press Close to exit the menu.
THE DEFAULT ALARM THRESHOLD IS 37.3 °C (99.14 °F).
WHEN MEASURING HIGH TEMPERATURE OBJECTS ENSURE THE
TEMPERATURE RANGE IS SET TO THE 0 °C TO 500 °C (32 °F TO
932 °F) RANGE TO OBTAIN THE MOST ACCURATE RESULTS.