8.Appendix
27
Appendix C: Fuse Replacement
Use the following procedure to replace the fuse.
1.
Ensure that the rotary switch is at the
position. Remove test leads and
any connectors from the input terminals.
2.
Lift the tilt stand and loosen the screws with a suitable Phillips screwdriver and
remove the battery cover.
3.
Pry the fuse cover open using a straight screwdriver, replace the fuse. Use only
the fuse of the specified rating (1 A / 250 V).
4.
Place the battery cover back in its original position and tighten the screws.