IceCure Medical Ltd.
DSR3210000 rev. D
ProSense™
Confidential
User Manual
99
Return to first page
5.3.8
Replace (refill) the Dewar during a cryoprocedure
Option to refill the Dewar is available at each THAW cycle.
In case you intend to do a long treatment that will freeze over 7 minutes in each
one of the freeze cycles, refill the Dewar during the THAW time of the
procedure.
First, always have an additional full Dewar ready, in case you may need it for a
longer freeze time.
•
At every THAW, the system will automatically ask if you want to replace the Dewar
(see figure below):
Figure 99: Replace Dewar option question
•
If you press CANCEL, you will stay in the THAW screen.
•
If you press OK you will get the
"Please wait… Do not open the door yet!"
message
meaning that the system is lowering the Dewar (see figure below).
•
If you don’t press either button, the message will disappear shor
tly before the end
of the THAW cycle.