Cryoprobe connection
Connect the cryoprobe to the cryohandle as follows, while maintaining sterility of the
probe:
•
Remove the plug that covers the probe connection point.
•
Insert the cryoprobe into the insertion point in the handle as shown on screen and
screw it until a NEXT button appears on screen then confirm screwing by an
additional slight rotation to confirm that probe connection is secured.
•
Remove the probe cover that protects the tip.
•
When done, press the Action button on the handle or NEXT on the touch screen.
•
After pressing NEXT
, a “Do not disconnect” message will appea
r
Figure 65: Cryoprobe connection screen
•
Probe reading failed
message appears if the probe reading failed. Make sure the
probe is secured and press OK to probe rereading.
Figure 66: Cryoprobe reading failed