IceCure Medical Ltd.
DSR3210000 rev. D
ProSense™
Confidential
User Manual
44
Return to first page
Figure 29: The 90 degrees cryohandle (No 1) and connected Cryoprobe (No 2)
Warning
Never reuse a single-use cryoprobe, temperature sensor or sterile sleeve.
Warning
At the conclusion of each patient session, remove and safely discard the
single-use cryoprobe, temperature sensor and sterile sleeve.
Caution
Artifacts from the devices may appear in the Ultrasound or CT image.
3.3.5.1
Handle plug holder and handle plug
The handle plug holder is located on the top of the main chassis and allows a place for
keeping the plug while performing the procedure.
Humidity is a crucial issue to address in cryosurgery systems.
After every procedure, close the handle plug on the Cryohandle.
Perform pretest at close proximity to the procedure treatment.
Figure 30: Handle plug holder and handle plug.