TECHNICAL INFORMATION
•
Type: TV01
•
Dimensions: 13 – 32 mm (W) * 45 – 80 mm (H) * 230 mm (L)
•
Weight: 155 g (without batteries), 250 g (with 4 x AA batteries)
•
Power supply: 4 x AA batteries
•
Measurement range: 1-‐99 mmHg
•
Accuracy of display: ±1
•
Measurement accuracy: ±2 (5-‐30 mmHg), 10% (30-‐80 mmHg)
•
Display unit: Millimetre mercury (mmHg)
•
The serial number is inside the battery compartment cover
•
There are no electrical connections from the tonometer to the patient
•
The device has B-‐type electrical shock protection
•
Operation environment:
o
Temperature: +10 °C to +35 °C
o
Relative humidity: 30% to 90%
o
Atmospheric pressure: 800 hPa – 1060 hPa
•
Storage environment:
o
Temperature: -‐10 °C to +55 °C
o
Relative humidity: 10% to 95%
o
Atmospheric pressure: 700 hPa – 1060 hPa
•
Transport environment:
o
Temperature: -‐40 °C to +70 °C
o
Relative humidity: 10% to 95%
o
Atmospheric pressure: 500 hPa – 1060 hPa
WARNING
No modification of this equipment is allowed
SYMBOLS
SPARE PARTS AND SUPPLIES
•
Probes in single-‐use package.
•
Probe base replacement kit.
Caution, Attention!!!
See instructions
Lot number
Do not dispose of in household
waste
Serial number
B-‐type device
See operating instructions for more
Single use only
Manufacturer
Temperature limitations
Keep dry