Icare   TONOVET   tonometer   TV01   manual            

   

 

   English   

 

 

 

www.icaretonometer.com

   

   

4   

TURNING   THE   TONOMETER   ON   

   
Insert   batteries   into   the   tonometer   (page   6).   
   

Place   the   wrist   strap   into   the   wrist   strap   attachment.   Place   the   wrist   strap   around   your   wrist.   The   wrist   strap   
protects   the   tonometer   from   dropping   down   accidentally.      
   
Press   the   measurement   button   to   turn   the   tonometer   ON.   The   tonometer   display   will   display   all   of   the   LCD   

segments   (see   the   image   beside).      
   
The   display   shows   “LoAd,”   instructing   the   user   to   load   the   single   use   probe   into   the   tonometer   prior   to   

measurement.   
   

   

LOADING   THE   PROBE   

   

Accurate   measurement   is   guaranteed   only   when   using   probes   made   by   Icare   
Finland   Oy.   
   
Open   the   probe   tube   by   removing   the   cap   and   insert   the   probe   into   probe   base   as   

shown   in   the   image.      
   
After   the   probe   has   been   inserted,   do   not   to   point   the   tonometer   down   to   prevent   

the   probe   from   falling   out.      
   
Activate   by   pressing   the   measurement   button   once.   The   display   shows   “00”   

instructing   that   the   tonometer   is   on   the   starting   position   and   ready   for   
measurement.      
   

After   activating   the   probe   is   magnetized   and   will   not   fall   out.   
   
   

   

ACCESSING   THE   STARTING   MODE   

   
Starting   mode   is   displayd   as   “00”.   
   

AFTER   MEASURING       

Press   the   measurement   button.   

BEFORE   LOADING   THE   PROBE   

Press   the   selector   button.   

   

   

MEASURING   THE   INTRAOCULAR   PRESSURE   

   

   
The   tonometer   needs   to   be   in   a   

horizontal   position

   during   the   measurement.   The   

distance   

should   be   4-­‐8   mm

   (1/6   -­‐   1/3   inch)   from   the   probe   to   the   cornea.   

   

TIP   

The   distance   from   the   eye   to   probe   should   be   about   the   lenght   of   the   collar.

   

   
Six   individual   measurements   are   required   for   the   final   result.   The   measurement   values   

displayed   during   the   measurement   represents   average   values   of   the   previous   measurements   
(1.-­‐5.),   not   individual   measurement   values.   
   

Press   the   measuring   button   six   times.   The   tip   of   the   probe   needs   to   contact   the   

central   

cornea

**.   After   each   successful,   individual   measurement   a   short   beep   is   played.   After   six   

successful   measurements,   a   longer   beep   is   played   and   the   final   result   is   shown   on   the   

display.   
   
If   there   is   an   erroneous   measurement,   the   tonometer   will   beep   twice   and   show   an   error   message.   Press   the   measurement   button   to   clear   the   error   
message   and   continue   the   sequence   of   six   measurements.   The   sequence   of   six   individual   measurements   does   not   have   to   be   started   from   beginning,   the   

tonometer   will   continue   the   sequence   after   cleared   error   message.   If   several   erroneous   measurements   appear,   see   the   section   ”Error   messages”   (page   7).   
   
Start   a   new   measurement   by   pressing   the   measurement   button.   The   display   shows   “00”   indicating   the   tonometer   is   ready   for   the   next   measurement.   

   
**IOP   should   always   be   measured   from   the   central   cornea   to   get   the   true   value   since   the   result   vary   in   different   parts   of   the   eye

   

 

   

Содержание Tonovet TV01

Страница 1: ...www icaretonometer com 1 ENGLISH USER S AND MAINTENANCE MANUAL...

Страница 2: ...ning the probe base 6 Cleaning the tonometer surface 6 Replacing the batteries 7 Error messages 7 Trouble shooting 7 Service procedures 8 Diagram of tonometer functions 8 Technical information 9 Symbo...

Страница 3: ...he substances that are flammable including flammable anaesthetic agents Check that a clean new disposable probe is used before each measurement To avoid transmitting certain microbiological agents e g...

Страница 4: ...EASURING THE INTRAOCULAR PRESSURE The tonometer needs to be in a horizontal position during the measurement The distance should be 4 8 mm 1 6 1 3 inch from the probe to the cornea TIP The distance fro...

Страница 5: ...1 8 2 5 Slight deviation the effect to the result is not likely to be relevant d Line in the middle the letter blinks 2 5 3 5 Deviation is greater than normal but the effect on the result is usually n...

Страница 6: ...ase every 12 months or if the cleaning is not effective CLEANING THE PROBE BASE The probe base should be cleaned every 6 months The probe base can be reused after careful cleaning Make sure the probe...

Страница 7: ...h a new one If the error message continues to appear replace the probe base E 02 The probe did not touch the eye The measurement was made from too far away or the probe base is dirty E 03 The probe sp...

Страница 8: ...e probe base if the probe does not move smoothly page 6 The device can be cleaned with a damp cloth containing disinfectant according the instructions page 6 There are no other service procedures that...

Страница 9: ...cal shock protection Operation environment o Temperature 10 C to 35 C o Relative humidity 30 to 90 o Atmospheric pressure 800 hPa 1060 hPa Storage environment o Temperature 10 C to 55 C o Relative hum...