7
English
Two short beeps. The battery is low.
Re-charge.
Two short beeps. The probe continuously
fails to move smoothly or
stops moving, because the
probe and/or probe base is
dirty, bent or twisted.
Follow the instructions in section 9,
press the main button to clear the
message and measure again.
•
Remove the probe base and the probe holder
from alcohol. Lay the probe holder on a
paper towel for a minimum of 5 minutes to
dry.
• Dry the probe base by blowing clean canned
or compressed air into the hole in the probe
base. This will additionally remove possible
residual dirt.
• Carefully without twisting, insert the dry
probe holder into the completely dry probe
base.
• Lay the tonometer on a table, the probe base
and forehead support facing up.
• Carefully without twisting, insert the probe
base, with the probe holder in it, onto the
tonometer.
• Carefully without twisting the probe base,
screw the collar back in, in order to lock the
probe base.
11. CLEANING THE TONOMETER
WARNING
Never spray, pour or spill liquid onto the Icare
tonometer, its accessories, connectors, switches
or openings in the chassis. Dry any liquid on the
surface of the tonometer immediately.
Icare PRO’s surfaces have been tested and
found chemically resistant to the following
liquids:
• 70-100 % isopropyl alcohol
• Mild soap solution
• 95% Pursept solution
Cleaning instructions for surfaces:
•
Turn the power off.
•
Dampen a soft cloth with one of the liquids
mentioned above.
• Lightly wipe the surfaces of the tonometer
with the soft cloth.
• Dry the surfaces with a dry soft cloth.
9. REPLACING THE PROBE BASE
Replace the probe base and probe holder every
six months. Replace or clean the probe base if
the tonometer displays the probe base error
‘Probe didn’t move properly’.
Instructions for replacing the probe base and
probe holder:
•
Turn off the tonometer.
• Unscrew the probe base collar and put it in a
safe place.
•
Remove the probe base and the probe holder
by tilting the tonometer downwards and use
your fingers to pull the probe base and the
probe holder out of the tonometer.
• Lay the tonometer on a table, the probe base
and forehead support facing up.
• Carefully without twisting, insert a new probe
base and probe holder into the tonometer.
• Carefully without twisting the probe base,
screw the collar back in, in order to lock the
probe base.
10. CLEANING THE PROBE BASE
You can reuse the probe base after careful
cleaning. Clean the probe base every three
months. Clean or replace the probe base if the
tonometer displays the probe base error ‘Probe
didn’t move properly’.
Instructions for cleaning the probe base:
• Fill the probe base cleaning container or
other clean container with 70-100% isopropyl
alcohol.
•
Turn the power off.
• Unscrew the probe base collar and carefully
remove the probe base.
•
Invert the probe base over the container,
drop in the probe base holder and the probe
base into the container and let soak for 5-30
minutes.
Содержание Pro TA03
Страница 2: ...3 4 5 6 3 7 mm 1 8 9 32 90 Icare tonometer PRO manual 1 2...
Страница 3: ...8 11 12 13 14 15 9 10 7...
Страница 17: ......
Страница 33: ......
Страница 75: ......
Страница 89: ......
Страница 103: ......
Страница 131: ......
Страница 174: ......
Страница 175: ......
Страница 176: ...www icaretonometer com...