7.4  TAKING THE MEASUREMENTS

 WARNING!

 

Do not bring the tonometer into contact with the eye or push it into the eye (the tip of the 

probe should be 4-8mm, or 5/32-5/16”, from the eye).

The probe will make a gentle and brief contact with the eye when you take the 

measurement. No topical anesthetic is needed.

To measure intraocular pressure: 

1.  Check that the Measure

 light (15) still flashes on the back panel.

2.  If the Measure

 light does not flash, press the 

power button (20) and wait until 

the Measure light illuminates again. 

3. 

The patient should look straight ahead at a specific point while keeping eyes 

wide open as shown in the below figure.

 

  

4.  Bring the tonometer near the eye, the probe pointing perpendicular to the 

center of the cornea without a vertical or horizontal tilt. The position is correct 

when the probe base light is green and appears symmetrically in the center of 

the patient’s view. See the below figures.

Correct head and eye position.

Correct measurement position.

Incorrect head and eye position.

Incorrect measurement position.

Incorrect head and eye position.

Incorrect measurement position.

5.  Press the measurement button:

Single mode: 

Press the button briefly (1 second) and you hear a 

short beep, repeat it to take one measurement at 

a time till you hear a long beep and see the Done 

light (17) illuminated on the back panel.

Series mode: 

Keep the measurement button down to obtain 

the sequence of six measurements till you hear a 

long beep and see the Done light illuminated on 

the back panel.

6.  If both eyes are measured repeat steps 1-5 using your other eye.
7. 

If an error occurs, press the Measure button briefly (1 second) and continue the 

measurement. See also section 9 Trouble shooting.

8. 

Press the power button for three seconds to turn the tonometer off.

9.  Dispose of the used probe.

Icare HOME tonometer Instruction manual for health care professionals

Icare HOME tonometer Instruction manual for health care professionals

English

English

90°

Correct measurement position.

Correct alignment of the 

tonometer.

Incorrect measurement position.

Incorrect alignment of the 

tonometer.

Incorrect measurement position.

Incorrect alignment of the 

tonometer. Readjust so that you see 

only the front of the tonometer and 

the green light symmetrically in the 

center of your view.

Содержание Icare HOME TA022

Страница 1: ...HOME INSTRUCTION MANUAL FOR HEALTH CARE PROFESSIONALS ENGLISH...

Страница 2: ...modification of this equipment is allowed WARNING Use only the original and certified probes made by the manufacturer The probes are for single use single pair of measurement sequences only Use probes...

Страница 3: ...arby devices for example sensitive sensors If you do not use the tonometer for a long time remove the batteries as they may leak Be sure to dispose of the single use probes properly for example in a c...

Страница 4: ...rly for damage to the case If you suspect that there is something wrong with the tonometer contact the dealer who sold the tonometer to you The package contains Icare HOME tonometer 10 sterilized sing...

Страница 5: ...re turned on briefly Following a brief pause the Load light flashes on the back panel to remind the user to load the single use probe into the tonometer prior to measurement 6 3 LOADING THE PROBE WARN...

Страница 6: ...ach of the six measurements to obtain the final IOP reading see figure left Series mode press the measurement button constantly Single mode press the measurement button briefly 1 second for each of th...

Страница 7: ...forehead and cheek supports with adjustment wheels 7 3 AUTOMATIC EYE RECOGNITION NOTE Do not cover the eye recognition transmitters or sensor during the measurement for example with your fingers Keep...

Страница 8: ...sition Incorrect measurement position Incorrect head and eye position Incorrect measurement position 5 Press the measurement button Single mode Press the button briefly 1 second and you hear a short b...

Страница 9: ...message a Set the correct distance 4 8mm 5 32 5 16 between the probe tip and the center of the cornea b Set the probe perpendicular to the center of the cornea Repeat light is flashing and Probe base...

Страница 10: ...At the same time the probe base light flashes red and you hear two short beeps The reason is either too much deviation in your measurement or the automatic eye recognition system could not recognize...

Страница 11: ...the tonometer Explain that the measurement button must be depressed to obtain the sequence of 6 measurements until a long beep is heard and the green Done light is illuminated on the back panel The pr...

Страница 12: ...certified prior to performing self tonometry STEP 1 Instruct the patient 1 Show and explain the tonometer s user interface including icons and status indicators Off no lights no signals Press power b...

Страница 13: ...strate measurements from your HCP s own eye 1 Position the tonometer on your HCP own eye as instructed above 2 Ask the patient to observe and learn STEP 3 Supervised patient use of the icare home 1 Lo...

Страница 14: ...strictions for professional use include Medivac vehicles or similar where vibration or noise levels are so high that the user cannot hear error signals Environmental restrictions for lay operators pat...

Страница 15: ...in all establishments including domestic establishments and those directly connected to public low voltage power supply network that supplies buildings used for domestic purposes Guidance and manufact...

Страница 16: ...he location in which the Icare HOME is used exceeds the applicable RF compliance level above the Icare HOME should be observed to verify normal operation If abnormal performance is observed additional...

Страница 17: ...www icaretonometer com 32 Icare HOME www icaretonometer com HOME Icare HOME tonometer Instruction manual for health care professionals English...

Отзывы: