background image

www.icaretonometer.com

www.icaretonometer.com

24

25

Icare HOME

Icare HOME

3.  The HCP loads the probe:

•  Unwraps the probe.
•  Removes the lid of the probe container.
•  Drops the probe into the probe base without touching the probe by 

holding the probe container.

•  Presses the measurement (play) button briefly (1 second) to activate 

the probe.

4.  Explain and show illustrations for how to position the tonometer (use a 

separate illustration sheet from the labeling for this):

•  Sit or stand in front of a mirror and hold the tonometer sideways in 

front of your face.

•  Align probe tip with center of cornea and rotate the tonometer until 

probe tip points straight at cornea.

•  Make sure probe base light is green. If probe base light is red, make 

sure you are facing straight ahead (i.e. head held at a 90° angle) and 

tilt tonometer until probe base light turns green.

•  The probe base light does not turn red in response to horizontal 

deviations as illustrated in the figure on the right. For this reason make 

sure the probe is centered in sight to ensure the probe contacts the 

center of cornea during measurement even if the probe base light is 

green. If the probe is not centered in your sight, repeat steps 5 and 6. 

This is very important because the tonometer with the probe must not 

be tilted more than 10 degrees away from the center of the cornea 

and without visualizing the probe base light it is difficult to judge the 

horizontal angle of the device.

5.  Explain how to take the measurement:

•  Explain that the Measure light will flash when the tonometer is ready 

to measure.

•  Explain that the user must take six individual measurements for the 

IOP result and that the results are stored in the tonometer.

•  Explain to the patient that the measurement button must be 

depressed to obtain the sequence of 6 measurements until a long 

beep is heard and the green “Done” light is illuminated on the back 

panel. The probe base light turns off at the same time.

STEP 2 – Demonstrate measurements from your (HCP’s) own eye

1.  Position the tonometer on your (HCP) own eye as instructed above.
2.  Ask the patient to observe and learn.

STEP 3 – Supervised patient use of the icare home

1.  Load a new probe and carefully, without touching the patient’s eye, choose the eye 

and set the distance between the tip of the probe and the center of the cornea for 

the patient at 4-8mm (5/32-5/16”) by turning knobs to adjust forehead and cheek 

support positioning as needed. Write the settings down on a support position tag for 

the patient. 

2.  Ask the patient to position the tonometer on the chosen eye and take some self-

measurements as instructed and demonstrated.

3.  Observe and, if necessary, correct the position while the patient positions the 

tonometer, say to the patient that this is the correct position and ask the patient to 

try again.

4.  Repeat 1-3 for up to 10 times until the patient shows consistent device positioning. 

If the patient cannot consistently measure after this point the patient will be 

considered to be unable to measure the IOP using the HOME device.

STEP 4 – Self-measurement by the patient

1.  Load a new probe and ask the patient to measure his/her IOP three times with  

 

the same Icare HOME tonometer.

2.   Observe if the positioning is correct. Do not supervise or interact.

STEP 5 –  Measurement of patient’s IOP by HCP

3.  The HCP measures the IOP of the patient once with the GAT tonometer.

STEP 6 – Certification

Connect device to computer and view the readings in the Icare LINK software.

The patient passes the training and is certified for self-use if the following conditions 

are met:

a.  The first of the three HOME readings taken by the patient and the GAT 

result measured by the HCP differ by 5 mmHg or less. 

b.  The range (max-min) of the three readings taken by the patient is 7 mmHg 

or less. 

c.  The positioning of the tonometer was correct during self-use as 

determined by the HCP.

Patient ID: 

 Date (dd mm yy): 

CERTIFICATION 

 0 = Pass; 1 = Fail  EYE 

 0 = Right; 1 = Left, 2 = Both

Icare HOME tonometer Instruction manual for health care professionals

Icare HOME tonometer Instruction manual for health care professionals

English

English

Содержание Icare HOME TA022

Страница 1: ...HOME INSTRUCTION MANUAL FOR HEALTH CARE PROFESSIONALS ENGLISH...

Страница 2: ...modification of this equipment is allowed WARNING Use only the original and certified probes made by the manufacturer The probes are for single use single pair of measurement sequences only Use probes...

Страница 3: ...arby devices for example sensitive sensors If you do not use the tonometer for a long time remove the batteries as they may leak Be sure to dispose of the single use probes properly for example in a c...

Страница 4: ...rly for damage to the case If you suspect that there is something wrong with the tonometer contact the dealer who sold the tonometer to you The package contains Icare HOME tonometer 10 sterilized sing...

Страница 5: ...re turned on briefly Following a brief pause the Load light flashes on the back panel to remind the user to load the single use probe into the tonometer prior to measurement 6 3 LOADING THE PROBE WARN...

Страница 6: ...ach of the six measurements to obtain the final IOP reading see figure left Series mode press the measurement button constantly Single mode press the measurement button briefly 1 second for each of th...

Страница 7: ...forehead and cheek supports with adjustment wheels 7 3 AUTOMATIC EYE RECOGNITION NOTE Do not cover the eye recognition transmitters or sensor during the measurement for example with your fingers Keep...

Страница 8: ...sition Incorrect measurement position Incorrect head and eye position Incorrect measurement position 5 Press the measurement button Single mode Press the button briefly 1 second and you hear a short b...

Страница 9: ...message a Set the correct distance 4 8mm 5 32 5 16 between the probe tip and the center of the cornea b Set the probe perpendicular to the center of the cornea Repeat light is flashing and Probe base...

Страница 10: ...At the same time the probe base light flashes red and you hear two short beeps The reason is either too much deviation in your measurement or the automatic eye recognition system could not recognize...

Страница 11: ...the tonometer Explain that the measurement button must be depressed to obtain the sequence of 6 measurements until a long beep is heard and the green Done light is illuminated on the back panel The pr...

Страница 12: ...certified prior to performing self tonometry STEP 1 Instruct the patient 1 Show and explain the tonometer s user interface including icons and status indicators Off no lights no signals Press power b...

Страница 13: ...strate measurements from your HCP s own eye 1 Position the tonometer on your HCP own eye as instructed above 2 Ask the patient to observe and learn STEP 3 Supervised patient use of the icare home 1 Lo...

Страница 14: ...strictions for professional use include Medivac vehicles or similar where vibration or noise levels are so high that the user cannot hear error signals Environmental restrictions for lay operators pat...

Страница 15: ...in all establishments including domestic establishments and those directly connected to public low voltage power supply network that supplies buildings used for domestic purposes Guidance and manufact...

Страница 16: ...he location in which the Icare HOME is used exceeds the applicable RF compliance level above the Icare HOME should be observed to verify normal operation If abnormal performance is observed additional...

Страница 17: ...www icaretonometer com 32 Icare HOME www icaretonometer com HOME Icare HOME tonometer Instruction manual for health care professionals English...

Отзывы: