Icare Icare HOME TA022 Скачать руководство пользователя страница 5

www.icaretonometer.com

www.icaretonometer.com

8

9

Icare HOME

Icare HOME

6. SETTING UP THE TONOMETER

Setting up your Icare HOME tonometer is easy, with few steps. The following 

subchapters describe how you get started.

6.1  INSTALLING OR CHANGING THE BATTERY

 NOTE!

Make sure you use batteries with built-in PTC protection, for example Energizer Lithium 

Photo 123 3V CR123A.
Update the tonometer’s time to your local time. It is done automatically by performing 

the steps 1 and 2 under section 8. Reading the measurement data.

Lift the silicon lid that protects the USB port and keeps 

the battery compartment cover in place. Open the 

battery compartment cover by pressing the silicon lid 

slightly and sliding the battery compartment cover as 

shown in the figure left.

1. Silicon lid 

2. Battery cover

Insert two CR123A lithium batteries in the correct 

order: (+) end upwards as shown in the figure left. 

Close the cover firmly and press the silicon lid in place 

to cover the USB port.

6.2  TURNING THE TONOMETER ON 

 

 NOTE!

The tonometer turns off automatically after 3 minutes if you do not use it.

Press the power button (20) to turn the tonometer on. The lights (14-19) are 

turned on briefly. Following a brief pause, the Load light flashes on the back 

panel to remind the user to load the single-use probe into the tonometer prior to 

measurement.

6.3  LOADING THE PROBE

 WARNING!

To prevent contamination, do not touch the bare probe, do not use a probe if it touches 

a non-sterile surface like a table or a floor.

 WARNING!

Use only the original and certified probes made by the manufacturer. The probes are for 

single-use (single pair of measurement sequences) only. Use probes taken only from the 

intact, original packaging. The manufacturer cannot guarantee sterility of the probe once 

the seal is compromised. Re-sterilization or re-use of the probe could result in incorrect 

measurement values, in the breakdown of the probe, cross-contamination of bacteria or 

viruses, and infection of the eye. Re-sterilization or re-use will void all responsibilities and 

liabilities of the manufacturer concerning the safety and effectiveness of the tonometer.

Icare HOME tonometer Instruction manual for health care professionals

Icare HOME tonometer Instruction manual for health care professionals

English

English

Opening the silicon lid and battery cover

Inserting new batteries

1.

2.

2 x CR123A

The Icare HOME tonometer uses single-use probes 

that are packed in a plastic tube and wrapped in blister 

packs as shown in the figures left.

Содержание Icare HOME TA022

Страница 1: ...HOME INSTRUCTION MANUAL FOR HEALTH CARE PROFESSIONALS ENGLISH...

Страница 2: ...modification of this equipment is allowed WARNING Use only the original and certified probes made by the manufacturer The probes are for single use single pair of measurement sequences only Use probes...

Страница 3: ...arby devices for example sensitive sensors If you do not use the tonometer for a long time remove the batteries as they may leak Be sure to dispose of the single use probes properly for example in a c...

Страница 4: ...rly for damage to the case If you suspect that there is something wrong with the tonometer contact the dealer who sold the tonometer to you The package contains Icare HOME tonometer 10 sterilized sing...

Страница 5: ...re turned on briefly Following a brief pause the Load light flashes on the back panel to remind the user to load the single use probe into the tonometer prior to measurement 6 3 LOADING THE PROBE WARN...

Страница 6: ...ach of the six measurements to obtain the final IOP reading see figure left Series mode press the measurement button constantly Single mode press the measurement button briefly 1 second for each of th...

Страница 7: ...forehead and cheek supports with adjustment wheels 7 3 AUTOMATIC EYE RECOGNITION NOTE Do not cover the eye recognition transmitters or sensor during the measurement for example with your fingers Keep...

Страница 8: ...sition Incorrect measurement position Incorrect head and eye position Incorrect measurement position 5 Press the measurement button Single mode Press the button briefly 1 second and you hear a short b...

Страница 9: ...message a Set the correct distance 4 8mm 5 32 5 16 between the probe tip and the center of the cornea b Set the probe perpendicular to the center of the cornea Repeat light is flashing and Probe base...

Страница 10: ...At the same time the probe base light flashes red and you hear two short beeps The reason is either too much deviation in your measurement or the automatic eye recognition system could not recognize...

Страница 11: ...the tonometer Explain that the measurement button must be depressed to obtain the sequence of 6 measurements until a long beep is heard and the green Done light is illuminated on the back panel The pr...

Страница 12: ...certified prior to performing self tonometry STEP 1 Instruct the patient 1 Show and explain the tonometer s user interface including icons and status indicators Off no lights no signals Press power b...

Страница 13: ...strate measurements from your HCP s own eye 1 Position the tonometer on your HCP own eye as instructed above 2 Ask the patient to observe and learn STEP 3 Supervised patient use of the icare home 1 Lo...

Страница 14: ...strictions for professional use include Medivac vehicles or similar where vibration or noise levels are so high that the user cannot hear error signals Environmental restrictions for lay operators pat...

Страница 15: ...in all establishments including domestic establishments and those directly connected to public low voltage power supply network that supplies buildings used for domestic purposes Guidance and manufact...

Страница 16: ...he location in which the Icare HOME is used exceeds the applicable RF compliance level above the Icare HOME should be observed to verify normal operation If abnormal performance is observed additional...

Страница 17: ...www icaretonometer com 32 Icare HOME www icaretonometer com HOME Icare HOME tonometer Instruction manual for health care professionals English...

Отзывы: