Do not install the device near washing basins, taps, etc. where it is likely to be splashed.
Do not lean on the door when it is opened.
Do not place trays, papers, fluid containers or other objects on the sterilizer.
Power supply connection
Check the label on the back panel of sterilizer to verify the voltage rating for the unit. Failure to
connect the sterilizer to an appropriate power supply could result in damage to the unit and
electrical shock to personnel.
Plug power cord into a properly polarized and grounded receptacle rated. A dedicated circuit only
used for the sterilizer is recommended. Never connect the device pin to reductions of any type.
Location requirements (mm)
4 Setup
Switch On and open the door to remove all of the inner contents for unpacking.
Connect the power cord to an outlet of the appropriate voltage.
Turn on the main power switch on the right side. After switching on, the machine turns on the LCD
and shows the door position, water level, working program, date, time, etc.
Fill the distilled water tank
Manual water filling
When the level of distilled water reaches a minimum level, the distilled water tank icon will flash
and beep three times.
Press the button on the tank lid and open it to the maximum position.
Fill it carefully with distilled water.
If exceeds the maximum level, an alarm will sound, and the distilled water tank icon will blink.