Drain the distilled water tank 

 

Attach the drain hose on a fitting connection located on the front of the sterilizer. Pull the 
connector to start the draining.  
Attention: The capacity of the distilled water tank is approximately 2.5 liters. 
 

Drain the used water tank 

 
Attach the drain hose on a quick fitting connection located inside the service door at the left.  
Attention: The capacity of the used water tank is approximately 2 liters. 

 

Preparation of sterilization materials 
 

For the most effective sterilization and to preserve the sample, please follow below: 
 

 

Clean instruments immediately after use. 

 

Treat the instruments by ultrasound cleaner. 

 

Residual chemicals left over after cleaning and disinfecting process may damage and 
corrode parts of the sterilizer, always rinse off the instruments using distilled water. 

 

Follow instrument manufacturer’s guidelines and recommendations for handing and 
cleaning instruments prior to sterilization. 

 

Check the manufacturer’s instructions as to proper procedure for sterilizing of each item. 

 

Arrange the samples of different materials on different trays or with at least 3cm of space 
between them. 

 

Clean and dry instruments thoroughly before placing them into tray. 

 

Always insert a sterilization paper or cloth between the tray and sample to avoid direct 
contact. 

 

Arrange the containers (glasses, cups, test-tubes, etc.) on one side or inverted position, 
avoiding possible water stagnation.  

 

Don’t stack the trays one above the other or put them in direct contact with the walls of 
the sterilization chamber. 

 

Always use the instrument tray handle. 

 

Wrap the samples one by one or, if more tools have to be set in the same bag, verify that 
these are made of the same material. 

 

Don’t use metallic clips, pins or other, as this jeopardizes the maintenance of the 
sterilizer. 

 

Don’t overload the trays over the stated limit (see appendix 2). 

 
 

 

Содержание STE-18-D

Страница 1: ...Instructions Manual For Models STE 18 D Pro STE 23 D Pro...

Страница 2: ...fully and follow all installation instructions Ningbo Ican Machines Co Ltd No 77 Yunlin East Road Gulin Town Ningbo China www icanclave com European Representative Icanclave Europe S L Juan Ramon Jime...

Страница 3: ...3...

Страница 4: ...ect the program 13 Start the sterilization program 14 End of cycle 14 Manually interruption of the cycle 15 Delay start cycle 15 Test Program 16 Descale 16 Data 17 Print and save report Optional 18 Pr...

Страница 5: ...only to a properly grounded outlet Do not cover or block any openings on this appliance Use this appliance only for its intended use as described in this manual Do not exceed the maximum weight limit...

Страница 6: ...door is open Fill the tank if it is flashing Min level of freshwater tank 0 5Liters Drain the tank if it is flashing The capacity is 1 5 liters 1 Report printer Grey No connected no paper Green Ready...

Страница 7: ...vent 3 Door lock 11 USB port 19 Condenser vent 4 Door handle 12 Main switch 20 Bacterial filter 5 Chamber 13 Distilled water tank vent 21 Used water tank vent optional 6 Door interlock 14 Used water t...

Страница 8: ...essure 76 kPa 106 kPa Packing content Item Accessories Quantity 1 Instruments tray 3 2 Instruments tray rack 1 3 Draining hose 2 4 Door Seal 1 5 Tray handle 1 6 Door adjustment wrench 1 7 Instructions...

Страница 9: ...s recommended Never connect the device pin to reductions of any type Location requirements mm 4 Setup Switch On and open the door to remove all of the inner contents for unpacking Connect the power co...

Страница 10: ...ument manufacturer s guidelines and recommendations for handing and cleaning instruments prior to sterilization Check the manufacturer s instructions as to proper procedure for sterilizing of each ite...

Страница 11: ...The Setup menu permits to set the following options Parameter Unit Preheat Expire date labels Water quality sensor Preheat Factory reset and see the information of Last error Select Setup from the mai...

Страница 12: ...um temperature to begin a sterilization cycle this helps to reduce the total cycle time and the drying efficiency The Preheat mode will be deactivated after one hour of inactivity Select Preheat from...

Страница 13: ...st it by pressing Press button to save and back Last Error In order to help the technical assistance process the most relevant information corresponding to the last error can be displayed on the scree...

Страница 14: ...n Select Program Select the program by pressing then press to confirm program in the screen will appear the information of selected program as the temperature and sterilization time holding time also...

Страница 15: ...Note If the cycle is manually interrupted before it reaches the drying phase the items inside the sterilizer must be considered not sterile N20 will appear on the screen see Error code description Ca...

Страница 16: ...perform the program automatically see appendix 2 After finishing the cycle you may check the indicator and evaluate the result according with the instructions of the test manufacturer B D test Put th...

Страница 17: ...ooled down Descale Mix the descaling powder with distilled water to make descaler according with the instruction of the descaling powder manufacturer From the Programs menu select descale After closin...

Страница 18: ...four numbers represent cycle number The last two numbers represent error code 00 no error 01 error E01 Sample of a printer report Program WRAPPED Temperature 134C Pressure 206 0 kPa Drying Time 08Min...

Страница 19: ...oor seal Select Settings from the main menu and tap Maintenance to enter in the menu Select the item to check the information Clean the distilled water tank Disconnect the main cable Drain the tank co...

Страница 20: ...you may adjust it Open the door insert your finger into the bottom of the door and pull the ring to unlock the mechanism Turn the lid counterclockwise to tighten it If it is too tight you may also tur...

Страница 21: ...t work The switch of locking system doesn t work E11 Failure to preheat the steam generator Power off run a new cycle Contact your supplier if error persists E12 Failure to preheat the chamber Power o...

Страница 22: ...tion Temporary interruption of winding 3 Safety valve Protection against possible sterilization chamber over pressure Action Release of the steam and restoration of the safety pressure 4 Safety micro...

Страница 23: ...oxide sio2 1mg I 1 0mg kg Iron 0 2mg I 0 1mg kg Cadmiun 0 005mg I 0 05mg kg Lead 0 05mg I 0 1mg kg Rest of heavy metals 0 1mg I 0 1mg kg Chloride 2mg I 0 1mg kg Phosphates 0 5 mg I 0 1mg kg Conductivi...

Страница 24: ...134 134 210 3 5 12 30 Unwrapped solid material 4 50 1 20 Wrapped 134 134 210 4 30 49 Unwrapped solid material 4 50 1 20 Single wrapped solid or hollow material 4 00 1 10 Wrapped 121 121 110 20 38 54...

Страница 25: ...id 134 134 210 4 15 30 Unwrapped solid material 5 00 1 50 Wrapped 134 134 210 4 38 57 Unwrapped solid material 5 00 1 50 Single wrapped solid or hollow material 4 50 1 20 Wrapped 121 121 110 20 46 62...

Отзывы: