Clean chamber, door seal ring, trays, and tray rack.
Remove the trays and tray rack from the chamber.
Clean trays, rack and inside of the chamber with mild soap.
Rinse the trays, rack and inside of the chamber with a smooth cloth and distilled water.
Examine door seal for possible damage.
Clean door seal and mating surfaces with a damp cloth.
Note: Do not use bleaching agents or any abrasive materials or substances in the chamber.
Failure to comply may cause damage to the chamber and/or other components.
Caution: To prevent burns, let the unit cool before cleaning gaskets and touch the
surface.
Door adjustment
Under normal circumstances, the chamber door does not require
adjustments. However, if the seal fails (resulting in steam leaking from the
front of the chamber), you may adjust it.
Open the door, insert your finger into the bottom of the door, and pull the
ring to unlock the mechanism. Turn the lid counterclockwise to tighten it. If it
is too tight, you may also turn the lid clockwise to loosen it.
Replacement of the door seal ring
Open the chamber door. Remove the door seal ring carefully by hand.
Clean the door seal ring carefully with a smooth cloth with distilled water.
Moisten the new seal with distilled water.
Insert the new seal and tap in sequence as follows:
Press in the top and bottom
of the door seal.
Press in the left and right
sides of the door seal.
Press the remaining sections
of the seal.
Caution: Please ensure the chamber and the door are cold prior to replacing the seal ring.
The drain valves
To drain the water from the used water tank or to fill the tank from an external container, insert the
connector attached to the silicone tube until a click is heard. To disconnect the connector, press
the metal button.