INTRODUCTION
INTRODUCTION
Thank you for selecting IC Sciences Corporation’s Blood Pressure Monitor
(LS810-B). The monitor features blood pressure measurement, pulse rate
measurement and the data storage. The design provides you with two
years of storage.
This manual contains
important
safety
and
care
informat
iion
and
provides step-by-step instructions for using the product.
Please read this user manual carefully and thoroughly before use.
FEATURES:
Systolic Blood Pressure
Diastolic Blood Pressure
Pulse Rate
Memory: Up to 60 records
This product uses the Oscillometric Measuring Method to detect blood pressure.
Before every measurement, the unit establishes a “zero point” equivalent to the
atmospheric pressure. Then it starts inflating the cuff. Meanwhile, the unit
detects pressure oscillation generated by beat-to-beat pulsatile, which is used
to determine the systolic pressure and diastolic pressure as well as pulse rate.
The device also compares the longest and the shortest intervals of detected
pulse wave with the average value, and then calculates the standard deviation.
The monitor will light up a warning symbol when the calculated standard
deviation is larger than or equal to 15.
Measurement Principle
General Description
The below signs might be in the user manual, labeling or other components.
They are the requirement of standard and using.
Safety Information
Symbol for “THE OPERATION
GUIDE MUST BE READ”
Symbol for “MANUFACTURER”
Symbol for “SERIAL NUMBER”
Symbol for “TYPE BF APPLIED
PARTS”
Symbol for “DIRECT CURRENT”
Symbol for “ENVIRONMENT
PROTECTION - Waste electrical products
should not be disposed of with
household waste. Please recycle where
facilities exist. Check with your local
authority or retailer for recycling advice”
The Bluetooth Combination Mark
Symbol for “MANUFACTURE
DATE”
2
3
Indications for use
The IC Sciences Corporation Blood Pressure Monitor is a digital
monitor intended for use in measuring blood pressure and heartbeat rate with
wrist circumference ranging from 13.5 cm to 21.5 cm ( about 5-8.5 inches ).
It is intended for adult indoor use only.
Symbol for “ Class II Equipment”
CAUTION
* It is intended for adult indoor use only. Pregnant women, pre-eclamptic patients,
and severely obese patients should not use the device. If you need to measure your blood
pressure, please consult your physician.
* This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is
not intended for use on extremities other than the wrist or for functions other than obtaining a
blood pressure measurement.
* Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood
pressure. Please start or end medical treatment basing solely on physician’s treatment advice.
* If you are taking medication, consult your physician to determine the most appropriate time for
your measurement. Never change a prescribed medication without your physician’s consent.
* This unit is not suitable for continuous monitoring during medical emergencies or operations.
Otherwise, the patient’s wrist and fingers will become anaesthetic, swollen, and even purple due
to a lack of blood.
* If the pressure of the cuff exceeds 40 kPa (300 mmHg), the unit will automatically deflate. If the
cuff doesn’t deflate when its pressure exceeds 40 kPa (300 mmHg), detach the cuff from the wrist
and press the START/STOP button to stop inflation.
* Do not use the monitor under the conditions of strong electromagnetic field (e.g. medical RF
equipment) that radiates interference signal or electrial fast transient/ burst signal.
* The maximum temperature that the applied part can be achieved is 42.5
℃
(108.5 F) while the
environmental temperature is 40
℃
(104 F)
.
* The device is not AP/APG equipment. It is not suitable for use in the presence of a flammable
anesthetic mixture with air (or oxygen, nitrous oxide).
* Please keep the unit out of reach of infants, children or pets, since inhalation or swallowing of
small parts is dangerous or even fatal.
* Please use ACCESSORIES and detachable parts specified / authorised by MANUFACTURER.
Otherwise, it may cause damage to the unit or danger to the user / patient.
* The patient is an intended operator. The patient can measure, transmit data and charge battery
under normal circumstances and maintain the device and its accessories according to the user
manual.
* The blood pressure monitor, and the cuff are suitable for use within the patient
environment. If you are allergic to dacron or plastic, please don’t use this device.
* The device is not intended for PATIENT transport outside a healthcare facility.
*This device cannot be used with HF surgical equipment at the same time.
* There is a PTC current limiter in the monitor, which specification is 8V and 0.5A. When the voltage
and current exceed the limiting value, the monitor will stop working.
* If Luer lock connectors are used in the construction of tubing, there is a possibility that they might
be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood
vessel.
* Be careful of strangulation due to cables and hoses, particularly due to excessive length.
Symbol for “Including RF
transmitter”
Readings taken by the LS810-B are equivalent to those obtained by a trained
observer using the cuff and stethoscope auscultation method.